FDA Adverse Event Injury Summary report: N

THE AFTERMARKET GROUP

MDR report key: 5952916 · Received September 7, 2016

Report

Report Number
5952916
Event Type
Injury
Date Received
September 7, 2016
Date of Event
September 4, 2016
Report Date
September 9, 2016
Manufacturer
UNITED PLASTIC MOLDERS, INC.
Product Code
KNO
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

SAFETY DEVICE THAT HOLDS THE NUT IN PLACE THAT CONNECTS BRAKES TO WHEELCHAIR DID NOT KEEP NUT IN PLACE DESPITE FOLLOWING MANUFACTURER'S INSTALLATION INSTRUCTIONS. POSITION OF BOLT ALLOWS TORQUE FROM BRAKES TO LOOSEN NUT BEYOND THE POSITIONING OF THE SAFETY CAP THAT COVERS THE NUT. WHEEL FELL OFF WHEELCHAIR CAUSING FALL WITH HEAD INJURY. ALL OF THESE SAME BRAKES ARE REMOVED FROM OTHER WHEELCHAIRS. UPON REMOVAL OF THESE SAME BRAKES FROM OTHER WHEELCHAIRS, THE NUT WAS LOOSE BEHIND EACH OF THE ORANGE SAFETY CAPS. IT APPEARS THAT THE ORANGE SAFETY CAP COVERS APPROXIMATELY 1/4 INCH OR LESS OF THE BOLT AND THE TORQUE FROM BRAKING RESULTS IN A LOOSENED NUT WITH THAT DOESN'T ENGAGE THE 1/4 INCH OF PROTECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
586313 THE AFTERMARKET GROUP SECURE AUTOMATIC WHEEL LOCK KIT KNO UNITED PLASTIC MOLDERS, INC.

Patients

Seq Age Sex Outcome Treatment
1 91 YR Hospitalization| L