STRATAMR VALVE, REGULAR
Report
- Report Number
- 2021898-2016-00321
- Event Type
- Malfunction
- Date Received
- September 14, 2016
- Date of Event
- August 16, 2016
- Report Date
- August 16, 2016
- Manufacturer
- MEDTRONIC NEUROSURGERY
- Product Code
- JXG
- PMA / PMN Number
- K152700
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE RETURNED VALVE WAS PATENT. THE VALVE MET THE REQUIREMENTS FOR SIPHON, AND REFLUX TESTING. ALL PERFORMANCE LEVELS COULD BE SET WITH A SINGLE ATTEMPT, AND THE VALVE DID NOT SPONTANEOUSLY CHANGE LEVELS DURING THE COURSE OF ANALYSIS. HOWEVER, THE DEVICE DID NOT MEET REQUIREMENTS FOR PRESSURE FLOW, AND PREIMPLANTATION TESTING. THERE WAS PROTEINACEOUS DEBRIS OBSERVED WITHIN THE INTERIOR AND EXTERIOR OF THE VALVE. DEBRIS WITHIN THE VALVE MAY HOLD PRESSURE-FLOW CONTROLLING MECHANISMS OPEN AFFECTING THE FLOW OF FLUID THROUGH THE VALVE. THE INSTRUCTIONS FOR USE THAT ACCOMPANY THE DEVICE CAUTION THAT ¿SHUNT OBSTRUCTION MAY OCCUR IN ANY OF THE COMPONENTS OF THE SHUNT SYSTEM. THE SYSTEM MAY BECOME OCCLUDED INTERNALLY DUE TO TISSUE FRAGMENTS, BLOOD CLOTS, TUMOR CELL AGGREGATES, BACTERIAL COLONIZATION, OR OTHER DEBRIS.¿ THE RETURNED VALVE DID NOT MEET THE REQUIREMENTS FOR LEAK TESTING DUE A TEAR OBSERVED IN THE TOP OF THE DELTA CHAMBER. THE INSTRUCTIONS FOR USE THAT ACCOMPANY THE DEVICE CAUTION THAT ¿CARE SHOULD BE TAKEN IN HANDLING THE VALVES AS SILICONE HAS A LOW CUT AND TEAR RESISTANCE.¿ IT IS ALSO SUGGESTED THAT THE VALVE BE PLACED IN A SURGICALLY CREATED LOOSE SUBGALEAL POCKET. THIS ALLOWS GENTLE PLACEMENT OF THE VALVE AND MINIMIZES HANDLING WITH SURGICAL TOOLS. A REVIEW OF THE MANUFACTURING RECORDS SHOWED NO ANOMALIES. ALL VALVES ARE 100% TESTED AT THE TIME OF MANUFACTURE.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE PRODUCT TESTING IS IN PROGRESS. THE INSTRUCTIONS FOR USE THAT ACCOMPANY THE DEVICE CAUTION THAT ¿CARE SHOULD BE TAKEN IN HANDLING THE VALVES AS SILICONE HAS A LOW CUT AND TEAR RESISTANCE.¿ IT IS ALSO SUGGESTED THAT THE VALVE BE PLACED IN A SURGICALLY CREATED LOOSE SUBGALEAL POCKET. THIS ALLOWS GENTLE PLACEMENT OF THE VALVE AND MINIMIZES HANDLING WITH SURGICAL TOOLS. A REVIEW OF THE MANUFACTURING RECORDS SHOWED NO ANOMALIES. ALL VALVES ARE 100% TESTED AT THE TIME OF MANUFACTURE.
IT WAS REPORTED TO MEDTRONIC NEUROSURGERY THAT THE DEVICE WAS FOUND TO BE LEAKING FROM THE DELTA CHAMBER INTRAOPERATIVELY. ACCORDING TO THE REPORT, THE DEVICE WAS REPLACED WITH ANOTHER DEVICE. REPORTEDLY, THERE WAS NO IMPACT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 602379 | STRATAMR VALVE, REGULAR | SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS | JXG | MEDTRONIC NEUROSURGERY | E12266 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR |