ARCHITECT C16000 SYSTEM
Report
- Report Number
- 1628664-2016-00232
- Event Type
- Malfunction
- Date Received
- September 13, 2016
- Date of Event
- August 24, 2016
- Report Date
- October 3, 2016
- Manufacturer
- ABBOTT MANUFACTURING INC
- Product Code
- JJE
- UDI-DI
- 00380740005924
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
AN ABBOTT FIELD SERVICE ENGINEER (FSE) ARRIVED AT THE CUSTOMER SITE AND INSPECTED THE ARCHITECT C16000 ANALYZER. THE FSE DETERMINED THE SOURCE OF THE LEAK BEING THE PERISTALTIC PUMP TUBING (LN 7-202464-01). THE CAUSE OF THE INSTRUMENT ERRORS WAS DETERMINED TO BE A DAMAGED AC/DC DRIVER BOARD (LN 7-202603-01), WHICH HAD EVIDENCE OF CHARRING AROUND ONE OF THE BOARD COMPONENTS. THE FSE REPLACED THE AC/DC DRIVER BOARD TO ADDRESS THE ERRORS BUT RETURNED THE FOLLOWING DAY AS THE ERRORS PERSISTED. IT WAS DETERMINED THAT THE NEW AC/DC DRIVER BOARD WAS DAMAGED AND HAD THE SAME APPEARANCE AS THE ORIGINAL BOARD. A THIRD BOARD WAS INSTALLED BUT ALSO FAILED IN A SIMILAR FASHION AS THE PREVIOUS TWO BOARDS. AFTER SIGNIFICANT TROUBLESHOOTING AND ELECTRICAL METERING DIFFERENT COMPONENTS ON THE SYSTEM, IT WAS DETERMINED THAT THE ORIGINAL LEAK FROM THE PERISTALTIC PUMP TUBING HAD DAMAGED THE VACUUM PUMP (LN 7-202560-01) LOCATED DIRECTLY BELOW. THE FSE DOCUMENTED THAT IT APPEARS THE DAMAGED VACUUM PUMP WAS DRAWING TOO MUCH CURRENT, WHICH THEN IN TURN LED TO DAMAGING THE AC/DC DRIVER BOARD. BOTH THE VACUUM PUMP AND THE AC/DC DRIVER BOARD WERE REPLACED TO RESOLVE THE ISSUE. NO RETURNS WERE REQUESTED FROM THE CUSTOMER SITE. PICTORIAL DOCUMENTATION PROVIDED BY THE FSE VERIFIED THE EVENT. SUBSEQUENT INSTRUMENT OPERATIONS WERE ACCEPTABLE. A REVIEW OF COMPLAINT TRACKING AND TRENDING METRICS WAS PERFORMED AND IDENTIFIED NO ADVERSE TRENDS IN CONJUNCTION WITH THE COMPLAINT ISSUE CURRENTLY UNDER EVALUATION. THE ARCHITECT SYSTEM OPERATIONS MANUAL AND THE ARCHITECT C16000 SYSTEM SERVICE AND SUPPORT MANUAL CONTAIN INFORMATION TO ADDRESS THE CURRENT CUSTOMER ISSUE. BASED ON THE AVAILABLE INFORMATION FROM THE CUSTOMER SITE AND FROM THE RESULTS OF THIS EVALUATION, THERE IS NO EVIDENCE TO REASONABLY SUGGEST A PRODUCT DEFICIENCY EXISTS. THE ISSUE WAS RESOLVED THROUGH STANDARD TROUBLESHOOTING PROCEDURES.
THE CUSTOMER CALLED TO REPORT A LEAK NEAR THE WASTE PUMP OF AN ARCHITECT C16000 ANALYZER. ALSO, THE CUVETTE WASHER WAS NOT RESPONDING AS EXPECTED AND THE CUSTOMER WAS UNABLE TO INITIALIZE THE INSTRUMENT INTO THE READY MODE. VARIOUS ERROR CODES (EC) WERE ALSO GENERATED: EC5623- ICT REFERENCE SOLUTION NOT ASPIRATED; EC3854- PROBE WASH PUMP TIMEOUT WHILE MOVING TO LOWER LIMIT; AND, EC5650 CUVETTE WASHER TIMEOUT WHILE MOVING TO UPPER LIMIT. AN ABBOTT FIELD SERVICE ENGINEER (FSE) VISITED THE CUSTOMER SITE AND REPLACED RUPTURED HIGH CONCENTRATION WASTE TUBING. THERE WERE NO INJURIES OR EXPOSURES DUE TO THIS ISSUE. SUBSEQUENT TROUBLESHOOTING LEAD TO THE DISCOVERY OF A DAMAGED AC/DC DRIVER BOARD. THE CUSTOMER LATER MENTIONED AN "ELECTRICAL FIRE" ODOR COMING FROM THE ANALYZER EARLIER IN THE DAY. THE FSE INSTALLED A NEW AC/DC CONTROLLER BOARD, WHICH DID NOT RESOLVE THE ISSUE. EXAMINATION OF THE BOARD FOUND CHARRING AROUND ONE OF THE COMPONENTS. THIS WAS LIMITED TO THE AC/DC DRIVER BOARD ONLY. THE FSE RETURNED TO THE SITE THE FOLLOWING DAY. DURING TROUBLE SHOOTING, TWO ADDITIONAL AC/DC DRIVER BOARDS FAILED. A NEW AC/DC DRIVER BOARD WAS INSTALLED ALONG WITH TWO VACUUM PUMPS, A POWER HARNESS AND A NEW WASH CUP MAG PUMP. SUBSEQUENT INSTRUMENT OPERATIONS WERE ACCEPTABLE. IN EACH CASE, CHARRING OF COMPONENTS ON EACH AC/DC DRIVER BOARD WAS OBSERVED. THERE WERE NO INJURIES REPORTED AND NO DAMAGE TO THE SURROUNDING LAB ENVIRONMENT. EACH INSTANCE WAS RESTRICTED TO THE RESPECTIVE AC/DC DRIVER BOARD. THERE IS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 597323 | ARCHITECT C16000 SYSTEM | AUTOMATED CHEMISTRY ANALYZER | JJE | ABBOTT MANUFACTURING INC | 00380740005924 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | BOARD, AC/DC DRIVER LN: (B)(4) |