FDA Adverse Event Malfunction Summary report: N

TRUEMETRIX

MDR report key: 5945529 · Received September 13, 2016

Report

Report Number
1000113657-2016-01501
Event Type
Malfunction
Date Received
September 13, 2016
Date of Event
August 18, 2016
Report Date
September 29, 2016
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT DOES NOT RETURN FOR EVALUATION YET. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE.

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND ON RETURNED METER AND RETURNED TEST STRIPS. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION:USER HAD AN INACCURATE REFERENCE.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 142, 124 AND 117MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 80 TO 110 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULTS OF 149 AND 133 MG/DL USING TRUE METRIX METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 10/21/2017 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 142, 124 AND 117MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 80 TO 110 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULTS OF 149 AND 133 MG/DL USING TRUE METRIX METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 10/21/2017 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
597152 TRUEMETRIX BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUEMETRIX MT1744

Patients

Seq Age Sex Outcome Treatment
1 0 YR SECOND THERAPY