FDA Adverse Event Other Summary report: N

BD PARADIGM LINK BLOOD GLUCOSE MONITOR

MDR report key: 594156 · Received June 11, 2004

Report

Report Number
2243072-2004-00276
Event Type
Other
Date Received
June 11, 2004
Date of Event
May 17, 2004
Report Date
June 10, 2004
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

CONSUMER CALLED FOR FAMILY MEMBER. BLOOD SUGAR TESTED TODAY WITH HIGH RESULTS, NURSE RETESTED WITH ANOTHER METER AND RESULTS WERE NORMAL. ANALYSIS RUN ON CLARK ERROR GRID USING COMPETITIVE READING (122) AS REFERENCE POINT VS BD READING (574) YIELDED DATA POINTS IN THE C RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD PARADIGM LINK BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 3078055

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other