PORTEX® BIVONA® ADULT TTS¿ TRACHEOSTOMY TUBES
Report
- Report Number
- 2183502-2016-01893
- Event Type
- Injury
- Date Received
- September 7, 2016
- Report Date
- August 10, 2016
- Manufacturer
- SMITHS MEDICAL ASD INC.
- Product Code
- JOH
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- RESPIRATORY THERAPIST
Narratives
THE 510(K): K913859 AND K083641. CUSTOMER HAS NOT RETURNED THE DEVICE TO THE MANUFACTURER FOR DEVICE EVALUATION. IF THE DEVICE BECOMES AVAILABLE AND IS RETURNED AND EVALUATED, THE MANUFACTURER WILL FILE A FOLLOW-UP REPORT DETAILING THE RESULTS OF THE EVALUATION.
THREE PORTEX® BIVONA® ADULT TTS¿ TRACHEOSTOMY TUBES WERE RETURNED FOR EVALUATION. IT COULD NOT BE DETERMINED WHICH DEVICE WAS ASSOCIATED WITH THE REPORTED EVENT; THEREFORE, THE EVALUATION OF ALL RETURNED DEVICES WILL BE USED FOR THE MEDWATCH. VISUAL INSPECTION (UNAIDED AND MICROSCOPICALLY) FOUND SCUFF MARKS NOTED BY AN ABSENCE OF AN EXPECTED COATING AT BOTH SIDES OF THE PROXIMAL END OF THE CUFF. FUNCTIONAL TESTING INVOLVED A LEAK TEST AND SHOWED THAT THE CUFF LEAKED FROM THE SCUFFED AREA ON THE AIRWAY SIDE OF EACH DEVICE. BASED ON THE EVIDENCE, THE ROOT CAUSE OF THE MARKS ON THE CUFFS WAS UNABLE TO BE DETERMINED. THERE WAS NO EVIDENCE FOUND TO SUGGEST THE EVENT WAS CAUSED FROM AN INTRINSIC DEFECT IN THE PRODUCT FROM THE MANUFACTURING PROCESS. (B)(4).
IT WAS REPORTED THAT A PORTEX® BIVONA® ADULT TTS¿ TRACHEOSTOMY TUBE LEAKED OVER THE PAST MONTH. REPORTER NOTED THAT FAMILY MEMBER HEARD AIR BLOWING OUT OF PATIENT'S MOUTH AND FOUND WATER LEAKING AT END OF CUFF. PATIENT LEFT THE PREVIOUS HOSPITAL VISIT WITH 9-10CC OF STERILE WATER IN THE CUFF AND FAMILY MEMBER CLAIMS 15CC WAS USED TO SEAL THE CUFF INTO THE TRACHEA. AN EMERGENCY TRACHEOSTOMY TUBE CHANGE WAS CONDUCTED AFTER ABOUT 10 DAYS DUE TO TUBE LEAKING. NO PERMANENT INJURY WAS REPORTED. SEE MFR: 2183502-2016-01894 AND 2183502-2016-01895.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 585576 | PORTEX® BIVONA® ADULT TTS¿ TRACHEOSTOMY TUBES | TUBE TRACHEOSTOMY AND TUBE CUFF | JOH | SMITHS MEDICAL ASD INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |