CATHETER MOUNT
Report
- Report Number
- 9611451-2016-00504
- Event Type
- Malfunction
- Date Received
- September 6, 2016
- Date of Event
- August 1, 2016
- Report Date
- August 8, 2016
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZO
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE RT021 CATHETER MOUNT IS SOLD IN THE USA BUT HAS NO 510(K) NUMBER AS IT IS CONSIDERED A CLASS I DEVICE. METHOD: THE COMPLAINT RT021 CATHETER MOUNT WAS RETURNED TO FPH IN (B)(4) FOR INVESTIGATION, WHERE IT WAS VISUALLY INSPECTED. OUR INVESTIGATION IS ALSO BASED ON OUR KNOWLEDGE OF THE PRODUCT AND INVESTIGATIONS OF PAST SIMILAR COMPLAINTS. RESULTS: VISUAL INSPECTION REVEALED THAT THE TUBING CUFF OF THE CATHETER MOUNT HAD SPLIT AT THE CONNECTOR END. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 160127. CONCLUSION: BASED ON THE INVESTIGATION CONDUCTED, THE SPLIT OBSERVED IS LIKELY DUE TO ENVIRONMENTAL STRESS CRACKING CAUSED BY NOT FULLY CURED PRIMER. ALL RT021 CATHETER MOUNTS ARE PRESSURE TESTED PRIOR TO RELEASE FOR DISTRIBUTION. ANY CATHETER MOUNT WITH A SPLIT TUBE WOULD HAVE FAILED THE PRESSURE TEST AND BE REJECTED FROM THE PRODUCTION LINE. OUR USER INSTRUCTIONS THAT ACCOMPANY THE RT021 CATHETER MOUNT STATE THE FOLLOWING: CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE. PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT
A MEDICAL CENTER IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE (FPH) REPRESENTATIVE THAT AN RT021 CATHETER MOUNT HAD A TEAR ON THE PLASTIC TUBING. THIS WAS DISCOVERED BEFORE PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 579761 | CATHETER MOUNT | BZO | BZO | FISHER & PAYKEL HEALTHCARE LTD | RT021 | 2100021460 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |