FDA Adverse Event Malfunction Summary report: N

ATTUNE FB TIBIAL IMPACTOR

MDR report key: 5924206 · Received September 2, 2016

Report

Report Number
1818910-2016-26282
Event Type
Malfunction
Date Received
September 2, 2016
Date of Event
August 23, 2016
Report Date
August 23, 2016
Manufacturer
DEPUY ORTHOPAEDICS, INC. 1818910
Product Code
LXH
PMA / PMN Number
K140881
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONFIRMED THAT THE IMPACTOR IS BROKEN AS REPORTED. EXPERT OPINION INDICATES THAT THE FAILURES ARE ASSOCIATED WITH ENVIRONMENTAL STRESS CRACKING (ESC). THE ATTUNE FB TIBIAL IMPACTOR HAS BEEN ANNEALED TO REDUCE RESIDUAL STRESSES. LAB TESTS INDICATE SIGNIFICANT REDUCTION IN RESIDUAL STRESSES OF ANNEALED SAMPLES IN COMPARISON TO UN-ANNEALED SAMPLES HOWEVER A COMPLETE REDUCTION IN STRESSES IS NOT ACHIEVABLE. THIS DEVICE IS FROM AN ANNEALED BATCH. A FIELD SAFETY NOTICE WAS ISSUED IN (B)(6) 2014 STATING TO REDUCE THE POSSIBILITY OF LEAVING FRAGMENTS IN PATIENTS, THE COMPANY SUGGESTS ADHERING TO THE INSTRUCTIONS FOR USE (IFU), WHICH INCLUDE INSPECTING THE IMPACTOR¿S TO ENSURE THAT NO INSTRUMENTS OR PIECES OF INSTRUMENTS ARE LEFT IN THE SURGICAL SITE PRIOR TO CLOSURE. IT SHOULD BE NOTED THAT CAPA-003835 HAS BEEN INITIATED AND CAN BE REFERENCED FOR FURTHER DETAILS. THE COMPLAINT SHALL BE CLOSED WITH A JUSTIFIED CONCLUSION; IT WILL BE ENTERED INTO THE COMPLAINT DATABASE AND MONITORED THROUGH TREND ANALYSIS. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.(B)(4).

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE IMPACTOR HAD A PIECE BREAK OFF WHILE IMPACTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
575950 ATTUNE FB TIBIAL IMPACTOR KNEE INSTRUMENTS LXH DEPUY ORTHOPAEDICS, INC. 1818910 AU3207735

Patients

Seq Age Sex Outcome Treatment
1 55 YR