PRAXAIR GRAB 'N GO III
Report
- Report Number
- 1000448358-2016-00003
- Event Type
- Malfunction
- Date Received
- August 31, 2016
- Date of Event
- July 28, 2016
- Report Date
- August 30, 2016
- Manufacturer
- PRAXAIR HEALTHCARE SERVICES (SPEC DEVELOPER)
- Product Code
- CAN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
BASED ON THE INVESTIGATIONS BY PRAXAIR IT HAS BEEN DETERMINED THIS UNIT WAS DAMAGED, AND THAT DAMAGE WAS NOT CONSISTENT WITH NORMAL WEAR AND TEAR. WE BELIEVE THIS IS AN ISOLATED INCIDENT BECAUSE OF THE FOLLOWING: A REVIEW OF THE PRAXAIR CUSTOMER COMPLAINTS AND NON CONFORMANCES WAS CONDUCTED AND THERE HAVE BEEN NO SIMILAR REPORTED INCIDENTS FOR GRAB 'N GO III MODELS OVER THE LAST YEAR. THE CUSTOMER ADMITTED TO NOT FOLLOWING OPERATING INSTRUCTIONS. THE GRAB 'N GO III UNITS ARE CHECKED FOR FLOW ON AN ANNUAL BASIS TO ENSURE NORMAL FUNCTIONING.
A REPORT WAS RECEIVED FROM (B)(6) HOSPITAL IN (B)(6) ON (B)(6) 2016 THAT A GRAB 'N GO III OXYGEN UNIT FAILED TO DELIVER A FLOW OF OXYGEN TO A PATIENT WHEN THE UNIT WAS TURNED ON. THE ADMINISTRATOR ACKNOWLEDGED THE DEVICE WAS CONNECTED TO THE PATIENT PRIOR TO PERFORMING THE VERIFICATION OF FLOW TEST, WHICH IS INCONSISTENT WITH THE INSTRUCTIONS IN THE USER MANUAL. THE UNIT WAS BEING USED ON A PATIENT THAT WAS IN TRANSPORT WITHIN THE HOSPITAL. THERE WAS NO INJURY TO THE PATIENT. ANOTHER GRAB 'N GO III OXYGEN UNIT WAS OBTAINED AND USED TO ADMINISTER OXYGEN TO THE PATIENT DURING TRANSPORT WITHOUT ISSUE. THE GRAB 'N GO III IS A CLASS I VALVE INTEGRATED PRESSURE REGULATING DEVICE, WHICH IS INSTALLED ONTO A STEEL HIGH PRESSURE MEDICAL OXYGEN USP CYLINDER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 569177 | PRAXAIR GRAB 'N GO III | REGULATOR, PRESSURE, GAS CYLINDER | CAN | PRAXAIR HEALTHCARE SERVICES (SPEC DEVELOPER) | PRX-9369 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |