ARCHITECT C8000 SYSTEM
Report
- Report Number
- 1628664-2016-00221
- Event Type
- Malfunction
- Date Received
- August 30, 2016
- Date of Event
- August 5, 2016
- Report Date
- September 13, 2016
- Manufacturer
- ABBOTT MANUFACTURING INC
- Product Code
- JJE
- UDI-DI
- 00380740000509
- PMA / PMN Number
- K980367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. AN EVALUATION IS IN PROCESS.
AN ABBOTT FIELD SERVICE REPRESENTATIVE (FSR) WAS DISPATCHED TO THE ACCOUNT AND FOUND THE ARCHITECT C8000 ANALYZER CENTRAL PROCESSING UNIT (CPU) WAS NOT COMMUNICATING WITH THE SYSTEM CONTROL CENTER (SCC). THE FSR REPLACED THE DATA ACQUISITION BOARD (DAQ) AND THE INSTRUMENT BOOTED UP SUCCESSFULLY. INSPECTION OF THE DAQ BOARD FOUND A DAMAGED CHIP. THE DISCOLORATION WAS ISOLATED TO A SINGLE CHIP AND THE SOLDER JOINTS FOR THIS CHIP ON THE BOARD. THE SMOKE FROM THE ANALYZER WAS LIMITED TO THIS PART AND DID NOT SPREAD TO OTHER PARTS OF THE EQUIPMENT. THE ANALYZER WAS RETURNED TO NORMAL OPERATION. EVALUATION OF THE CUSTOMER ISSUE INCLUDED A REVIEW OF THE COMPLAINT TEXT, A SEARCH FOR SIMILAR COMPLAINTS, A REVIEW OF LABELING, AND A REVIEW OF INSTRUMENT SERVICE. RETURNS WERE MADE AVAILABLE FROM THE CUSTOMER SITE FOR THIS EVALUATION, IN ADDITION TO PHOTOGRAPHS OF THE DAMAGED DAQ BOARD. ARCHITECT C8000 ANALYZER SERIAL NUMBER (B)(4) SERVICE HISTORY WAS REVIEWED AND NO CONTRIBUTING FACTORS WERE FOUND. NO SUBSEQUENT REPORTS OF SMOKE/BURN ISSUES WERE IDENTIFIED. HISTORICAL COMPLAINT DATA WAS REVIEWED FOR A 62 MONTH PERIOD AND NO ADVERSE TREND WAS IDENTIFIED. NO SIMILAR COMPLAINTS ASSOCIATED WITH THE DAQ BOARD WERE FOUND. PRODUCT LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. THE ISSUE WAS RESOLVED THROUGH NORMAL TROUBLESHOOTING PROCEDURES. BASED ON ALL AVAILABLE INFORMATION AND ABBOTT DIAGNOSTICS COMPLAINT INVESTIGATION FOR THE ARCHITECT C8000 ANALYZER, NO PRODUCT DEFICIENCY WAS IDENTIFIED.
THE CUSTOMER OBSERVED A BURNING ODOR FROM THE BACK LEFT SIDE OF THE ARCHITECT C8000 ANALYZER IN THE CARD CAGE AREA. THE CUSTOMER WAS RUNNING QC AT THE TIME OF THE ISSUE, AND NO PATIENT SAMPLES WERE IMPACTED. THE CENTRAL PROCESSING UNIT (CPU) FAILED TO BOOT UP AND DID NOT CONNECT TO THE SYSTEM CONTROL CENTER (SCC). ABBOTT FIELD SERVICE REPRESENTATIVE (FSR) REPLACED THE DATA ACQUISITION BOARD (DAQ). THE FSR OBSERVED EVIDENCE OF A DAMAGED CHIP ON THE DAQ THAT HAD OVERHEATED. NO FIRE OR SMOKE WAS REPORTED AND NO INJURIES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 566741 | ARCHITECT C8000 SYSTEM | AUTOMATED CHEMISTRY ANALYZER | JJE | ABBOTT MANUFACTURING INC | 00380740000509 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |