FDA Adverse Event Malfunction Summary report: N

EXACT 47

MDR report key: 590841 · Received April 8, 2005

Report

Report Number
1034973-2005-00001
Event Type
Malfunction
Date Received
April 8, 2005
Date of Event
March 10, 2005
Report Date
March 10, 2005
Manufacturer
CTI PET SYSTEMS, INC.
Product Code
KPS
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

DURING ROUTINE PREVENTATIVE MAINTENANCE ON A PET MOBILE COACH, THE SERVICE ENGINEER DISCOVERED INTERNAL WEAR AND TWO BROKEN SCREWS IN THE VERTICAL MOTOR MOUNT, WHICH IS AT THE BASE OF THE PT HANDLING SYSTEM (PHS). CPS HAS INVESTIGATED THIS PROBLEM AND FOUND THAT THE SUPPORT SCREWS ON TEH PHS VERTICAL MOTOR MOUNT CAN BECOME LOOSE OVER TIME DUE TO INCREASED VIBRATION. THIS PROCESS THEN CAUSES STRESS AND BREAKAGE IN OTHER MOUNTING SCREWS. THIS WEAR AND BREAKAGE COULD POTENTIALLY CAUSE THE PHS TO FALL. A BED SUPPORT IS REQUIRED WHILE THE SYSTEM IS IN TRANSIT. THIS REDUCES STRESS AND VIBRATION ON THE PHS WHILE THE MOBILE COACH IS IN TRANSIT. IT IS UNK WHETHER THIS CUSTOMER HAD PROPERLY UTILIZED THE BED SUPPORT, WHICH COULD HAVE CONTRIBUTED TO THE PROBLEM. NO INJURIES HAVE BEEN REPORTED TO CPS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXACT 47 PET SCANNER KPS CTI PET SYSTEMS, INC. 3600088-00 NA

Patients

Seq Age Sex Outcome Treatment
1 NA Other