EXACT 47
Report
- Report Number
- 1034973-2005-00001
- Event Type
- Malfunction
- Date Received
- April 8, 2005
- Date of Event
- March 10, 2005
- Report Date
- March 10, 2005
- Manufacturer
- CTI PET SYSTEMS, INC.
- Product Code
- KPS
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
DURING ROUTINE PREVENTATIVE MAINTENANCE ON A PET MOBILE COACH, THE SERVICE ENGINEER DISCOVERED INTERNAL WEAR AND TWO BROKEN SCREWS IN THE VERTICAL MOTOR MOUNT, WHICH IS AT THE BASE OF THE PT HANDLING SYSTEM (PHS). CPS HAS INVESTIGATED THIS PROBLEM AND FOUND THAT THE SUPPORT SCREWS ON TEH PHS VERTICAL MOTOR MOUNT CAN BECOME LOOSE OVER TIME DUE TO INCREASED VIBRATION. THIS PROCESS THEN CAUSES STRESS AND BREAKAGE IN OTHER MOUNTING SCREWS. THIS WEAR AND BREAKAGE COULD POTENTIALLY CAUSE THE PHS TO FALL. A BED SUPPORT IS REQUIRED WHILE THE SYSTEM IS IN TRANSIT. THIS REDUCES STRESS AND VIBRATION ON THE PHS WHILE THE MOBILE COACH IS IN TRANSIT. IT IS UNK WHETHER THIS CUSTOMER HAD PROPERLY UTILIZED THE BED SUPPORT, WHICH COULD HAVE CONTRIBUTED TO THE PROBLEM. NO INJURIES HAVE BEEN REPORTED TO CPS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EXACT 47 | PET SCANNER | KPS | CTI PET SYSTEMS, INC. | 3600088-00 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |