CAD II
Report
- Report Number
- 8030965-2016-14633
- Event Type
- Malfunction
- Date Received
- August 24, 2016
- Report Date
- July 22, 2016
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
(B)(4). (B)(6). THIS DEVICE WAS RETURNED FOR SERVICE; HOWEVER, DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS CONFIRMED. HOWEVER, THE ASSIGNABLE ROOT CAUSE COULD NOT BE DETERMINED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY.
IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS OBSERVED THAT THE COUPLING POWER SUPPLY WAS DEFORMED, BENT AND THE COUPLING SHAFT WAS BROKEN ON THE COMPACT AIR DRIVE DEVICE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED THE FOLLOWING PRE-TESTS: GENERAL CONDITION, CHECK THE ATTACHMENT COUPLING, CHECK ATTACHMENT COUPLING WITH ATTACHMENTS, CHECK REVERSE LOCKING MECHANISM, CHECK AIR HOSE COUPLING, CHECK FOR AIR LEAK, CHECK FUNCTION OF SOFT MODE SWITCH (SAFETY SYSTEM), CHECK FOR UNTRUE RUNNING, CHECK FOR EXCESSIVE NOISE, CHECK THE POWER WITH TEST BENCH AND CHECK STARTING BEHAVIOR. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE WAS BLOCKED. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 551206 | CAD II | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT | HWE | DEPUY SYNTHES POWER TOOLS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |