Report

Report Number
8030965-2016-14633
Event Type
Malfunction
Date Received
August 24, 2016
Report Date
July 22, 2016
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). THIS DEVICE WAS RETURNED FOR SERVICE; HOWEVER, DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS CONFIRMED. HOWEVER, THE ASSIGNABLE ROOT CAUSE COULD NOT BE DETERMINED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS OBSERVED THAT THE COUPLING POWER SUPPLY WAS DEFORMED, BENT AND THE COUPLING SHAFT WAS BROKEN ON THE COMPACT AIR DRIVE DEVICE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED THE FOLLOWING PRE-TESTS: GENERAL CONDITION, CHECK THE ATTACHMENT COUPLING, CHECK ATTACHMENT COUPLING WITH ATTACHMENTS, CHECK REVERSE LOCKING MECHANISM, CHECK AIR HOSE COUPLING, CHECK FOR AIR LEAK, CHECK FUNCTION OF SOFT MODE SWITCH (SAFETY SYSTEM), CHECK FOR UNTRUE RUNNING, CHECK FOR EXCESSIVE NOISE, CHECK THE POWER WITH TEST BENCH AND CHECK STARTING BEHAVIOR. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE WAS BLOCKED. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
551206 CAD II INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT HWE DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1