TRUEMETRIX
Report
- Report Number
- 1000113657-2016-01396
- Event Type
- Malfunction
- Date Received
- August 23, 2016
- Date of Event
- August 1, 2016
- Report Date
- August 23, 2016
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- PMA / PMN Number
- K140100
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM
- Reporter Occupation
- OTHER
Narratives
(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECT FOUND ON RETURNED METER. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER'S TEST STRIP HAD A POOR STORAGE(KITCHEN).
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 139, 152, 163, 159MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 90 TO 125MG/DL FASTING AND WITHIN TWO HOURS AFTER EATING BETWEEN 160-180 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER ON-FASTING AND PRODUCED TEST RESULTS OF 141 AND 136 MG/DL USING TRUE METRIX METER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION,CUSTOMER STORES PRODUCT IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 07/31/2017 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: 139MG/DL (B)(6) 2016 07:35 AM FASTING: YES, 152MG/DL (B)(6) 2016 07:33 AM FASTING: YES, 163MG/DL (B)(6) 2016 07:32 AM FASTING: YES, 159MG/DL (B)(6) 2016 07:31 AM FASTING: YES, 121MG/DL (B)(6) 2016 07:28 AM FASTING: YES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 548388 | TRUEMETRIX | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | TRUEMETRIX | MT1575 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |