FDA Adverse Event Injury Summary report: N

COOK CELECT® PLATINUM NAVALIGN UNISET VENA CAVA FILTER SET

MDR report key: 5895776 · Received August 23, 2016

Report

Report Number
3002808486-2016-00939
Event Type
Injury
Date Received
August 23, 2016
Date of Event
June 5, 2015
Report Date
July 9, 2015
Product Code
DTK
UDI-DI
10827002345048
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GB
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CATALOG #: IGTCFS-65-2-UNI-CELECT-PT. SIMILAR TO DEVICE UNDER 510(K) K121629. SUMMARY OF INVESTIGATIONAL FINDINGS: BASED ON THE IMAGING REVIEW BILATERAL PE WAS DEMONSTRATED, BUT THE REPORTED DVT ~66 DAYS AFTER FILTER PLACEMENT COULD NOT BE CONFIRMED, AS REFERENCED ULTRASOUND WAS NOT PROVIDED. LT IS POSSIBLE BUT VERY UNLIKELY THE RIGHT LEG DVT WAS A RIGHT CFV ACCESS SITE COMPLICATION GIVEN THE NORMAL CFV ON ULTRASOUND PRE-IMPLANTATION AND THE LACK OF SYMPTOMS FOR TWO MONTHS FOLLOWING IMPLANTATION. THERE IS NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. COOK MEDICAL WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS.

Description of Event or Problem · 1

DESCRIPTION OF EVENT ACCORDING TO STUDY: ON (B)(6) 2015 (94 DAY POST-PROCEDURE), AN UNSCHEDULED ULTRASOUND WAS COMPLETED. THE ULTRASOUND REVEALED NO THROMBUS IN THE INFERIOR VENA CAVA (IVC) OR THE PELVIC VEINS. THERE WAS THROMBUS IN THE RIGHT LOWER EXTREMITIES. THIS IMAGING RESULT WAS ASSOCIATED WITH CLINICAL SEQUELAE AND LED TO AN INTERVENTION OR TREATMENT. THE EVENT WAS A RIGHT DEEP VEIN THROMBOSIS (DVT) FIRST DIAGNOSED ON (B)(6) 2015 (66 DAY POST-PROCEDURE) . THE PATIENT WAS TREATED WITH ANTICOAGULATION, WARFARIN AND DALTEPARIN. PATIENT OUTCOME: THE PATIENT REMAINS IN THE STUDY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
548481 COOK CELECT® PLATINUM NAVALIGN UNISET VENA CAVA FILTER SET DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK 10827002345048

Patients

Seq Age Sex Outcome Treatment
1 69 YR Life Threatening