QUANTA-LYSER
Report
- Report Number
- 2026994-2016-00001
- Event Type
- Injury
- Date Received
- August 22, 2016
- Date of Event
- July 25, 2016
- Report Date
- August 22, 2016
- Manufacturer
- INOVA DIAGNOSTICS, INC.
- Product Code
- JJE
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION SUMMARY: ON 8-9-2016, INOVA FIELD SERVICE TECH. CONFIRMED THAT COVER SENSOR WAS NOT ACTIVATED AT THE USER'S SITE. THE INSTRUMENT WAS NOT RETURNED TO INOVA. THE USER'S MANUAL CONTAINS MULTIPLE WARNINGS REGARDING USE OF THE INSTRUMENT WHILE IN OPERATION. THERE ARE NO EXPLICIT WARNING SIGNS ON THE INSTRUMENT REGARDING THE USE OF THE INSTRUMENT WHILE IN OPERATION THAT ARE UTILIZED. ON 8-9-2016, INOVA DIAGNOSTICS FIELD SERVICE TECHNICIAN ACTIVATED THE COVER SENSOR ON THE QUANTA-LYSER, S/N (B)(4), AT THE USER'S SITE TO PREVENT OPERATION IF THE COVER IS OPEN.
THE CUSTOMER REPORTED THAT THEIR EMPLOYEE WAS LOADING THE QUANTA-LYSER INSTRUMENT WHILE IT WAS RUNNING AND HAD THE HAND IMPALED BY THE PIPETTES. THE TECHNICIAN THAT SUFFERED THE INJURY HAD ALL FOUR SAMPLE PROBES PIERCE HER HAND. SHE HAS BEEN TO THE HOSPITAL BUT ADMISSION STATUS IS UNKNOWN. SHE HAS BEEN PRESCRIBED ANTI-RETROVIRAL MEDICINE AS A PRECAUTION. BLOODBORNE PATHOGEN EXPOSURE INFORMATION IS PENDING. THE QUANTA-LYSER IS AN AUTOMATED BENCH TOP BATCH PROCESSOR DESIGNED FOR PERFORMING AUTOMATICALLY VARIOUS STEPS SUCH AS PIPETTING, MEASURING COLOR INTENSITY, ETC. FOR USE WITH ENZYME IMMUNOASSAYS, ELISA AND IMMUNOFLUORESCENT ASSAY TECHNOLOGIES. THE QUANTA-LYSER HAS A BUILT-IN CAPABILITY TO PREVENT OPERATION IF THE FRONT COVER IS OPEN. IT IS REFERENCED IN THE USER MANUAL IN THE PRECAUTIONS SECTION. THE COVER SENSOR FUNCTIONALITY IS INACTIVE BY DEFAULT AND MUST BE ACTIVATED BY AN INOVA FIELD SERVICE TECHNICIAN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 544554 | QUANTA-LYSER | CHEMISTRY ANALYZER FOR CLINICAL USE | JJE | INOVA DIAGNOSTICS, INC. | 240 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |