BLACKMAX-NEURO
Report
- Report Number
- 1045834-2016-12489
- Event Type
- Malfunction
- Date Received
- August 19, 2016
- Report Date
- July 19, 2016
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HSZ
- PMA / PMN Number
- K831756
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
THE DATE OF MANUFACTURE WAS DOCUMENTED AS UNKNOWN IN THE INITIAL REPORT. IT HAS BEEN UPDATED AS JUN 25, 1995. (B)(4). THIS DEVICE WAS RETURNED FOR SERVICE; HOWEVER, DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS DUPLICATED AND CONFIRMED. IT WAS ALSO OBSERVED THAT THE DEVICE POWER WAS LOW. THE DEVICE FAILED THE FOLLOWING PRE-TESTS: TEMPERATURE, INITIAL ROTATIONAL AND NOISE VERIFICATION. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO COMPONENT DAMAGE FROM NORMAL USE AND SERVICING OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY.
ADDITIONAL NARRATIVE: THE DATE OF MANUFACTURE IS UNKNOWN. AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED IN THE SERVICE ORDER FROM (B)(6) THAT THE MOTOR DEVICE WAS OVERHEATING. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 541891 | BLACKMAX-NEURO | MOTOR, DRILL, PNEUMATIC - HANDPIECE | HSZ | DEPUY SYNTHES POWER TOOLS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |