FDA Adverse Event Malfunction Summary report: N

CAD II

MDR report key: 5886269 · Received August 18, 2016

Report

Report Number
8030965-2016-14576
Event Type
Malfunction
Date Received
August 18, 2016
Report Date
August 5, 2016
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). THIS DEVICE WAS RETURNED FOR SERVICE; HOWEVER, DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. (B)(4) EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO IMPROPER HANDLING, WHICH IS USER ERROR/MISUSE/ABUSE. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS OBSERVED THAT THE COUPLING POWER SUPPLY ON THE COMPACT AIR DRIVE DEVICE WAS DEFORMED AND BENT. IT WAS FURTHER DETERMINED THAT THE MOTOR HAD DAMAGE AND DEFORMATION IN THE INTERRUPTER AND GENERAL WEAR AND TEAR OF INTERNAL COMPONENTS. THE DEVICE ALSO FAILED THE FOLLOWING PRE-TESTS: CHECK STATUS OF DEVELOPMENT, GENERAL CONDITION, CHECK AIR HOSE COUPLING, CHECK FOR AIR LEAK, CHECK FOR UNTRUE RUNNING, CHECK FOR EXCESSIVE NOISE AND CHECK STARTING BEHAVIOR. IT WAS ALSO REPORTED IN THE SERVICE ORDER THAT THE DEVICE HAD DAMAGE AND DEFORMATION IN THE INTERRUPTER OF THE EQUIPMENT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
537032 CAD II INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT HWE DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1