FDA Adverse Event Malfunction Summary report: N

TRINITY

MDR report key: 5881729 · Received August 17, 2016

Report

Report Number
9614209-2016-00135
Event Type
Malfunction
Date Received
August 17, 2016
Report Date
August 17, 2016
Manufacturer
CORIN MEDICAL
Product Code
LWJ
PMA / PMN Number
K093472
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

C-1592 INITIAL REPORT. THIS INITIAL REPORT IS BEING SUBMITTED NOW AS IT WAS ERRONEOUSLY SUBMITTED AS A FOLLOW-UP REPORT IN ERROR. THIS CASE WAS SUBMITTED AS A RESULT OF A RETROSPECTIVE REVIEW. NO ADVERSE EVENT REPORTED. ADDITIONAL INFORMATION, INCLUDING THE RETURN OF THE DEVICE WAS REQUESTED IN ORDER TO PROGRESS WITH THE INVESTIGATION, BUT WAS NOT RECEIVED. THEREFORE, THERE WAS ONLY LIMITED INFORMATION AVAILABLE FOR THIS INVESTIGATION. THE APPROPRIATE DEVICE DETAILS WERE PROVIDED AND THE RELEVANT DEVICE MANUFACTURING RECORDS WERE IDENTIFIED AND REVIEWED. UNFORTUNATELY THE REPORTED DEVICE WAS NOT RETURNED TO CORIN UK FOR EXAMINATION, HOWEVER, THE POSSIBLE ROOT CAUSES OF THE FAILURE MODE WERE DISCUSSED AND IT WAS CONCLUDED THAT THE THREAD ON THE HANDLE MAY HAVE BEEN DAMAGED PRIOR TO ATTACHMENT OF THE CUP WHICH MAY HAVE LED TO CROSS THREADING. BASED ON THIS, CORIN NOW CONSIDERS THIS CASE CLOSED. HOWEVER, SHOULD ANY NEW INFORMATION BE PROVIDED OR IF THE DEVICE IS RETURNED THIS CASE MAY BE RE-OPENED FOR FURTHER INVESTIGATION. PLEASE NOTE: THIS REPORT IS FILED WITH THE FDA DUE TO AN ADVERSE EVENT EXPERIENCED WITH A DEVICE THAT IS SIMILAR TO THOSE PLACED ON THE MARKET IN THE USA. THIS EVENT OCCURRED OUTSIDE OF THE USA.

Description of Event or Problem · 1

THE TRINITY STD INTRODUCER/IMPACTOR HANDLE COULD NOT BE REMOVED FROM THE CUP AFTER CUP IMPACTION DUE TO THE BLACK ACETAL HANDLE ROTATING AROUND THE SHAFT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
536745 TRINITY ACETABULAR HIP SYSTEM LWJ CORIN MEDICAL 921.129 059814-51

Patients

Seq Age Sex Outcome Treatment
1 Unknown