TRUBALANCE
Report
- Report Number
- 1000113657-2016-01350
- Event Type
- Malfunction
- Date Received
- August 15, 2016
- Date of Event
- July 23, 2016
- Report Date
- August 15, 2016
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- PMA / PMN Number
- K090495
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS
- Reporter Occupation
- OTHER
Narratives
(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECT FOUND ON RETURNED METER. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE.
CONSUMER REPORTED COMPLAINT FOR ERRATIC BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM BACK TO BACK BLOOD TEST OF 254 AND 340MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 114-170 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER NON-FASTING AND PRODUCED TEST RESULTS OF 156 AND 192MG/DL. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 8/31/2018 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6). MEMORY CONCERNS: THE CUSTOMER IS CONCERNED WITH THE RESULT OF 231, 187, 340 AND 254MG/DL IN THE METER'S MEMORY. THE CUSTOMER STATED THAT IS TESTING ON FASTING MODE. THE CUSTOMER IS CONCERNED THAT DID THE TESTS A FEW MINUTES APART AND THE RESULTS AND THE VARIANCE WAS WIDE. MANUFACTURER RECOMMENDED NO TEST IMMEDIATELY AFTER ONE FIRST TIME. THE CUSTOMER HAS NOT HAD ANY CHANGES IN THE DIET. CUSTOMER EXPRESSED CONCERN WITH THE BACK TO BACK BLOOD TEST THAT WAS TAKEN. CUSTOMER EDUCATED ABOUT THE ACCEPTABLE VARIANCE WITH A BACK TO BACK TEST IS 20%.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 529350 | TRUBALANCE | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | TRUBALANCE | BT4885 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |