TEMPOROMANDIBULAR JOINT REPLACEMENT SYSTEM
Report
- Report Number
- 0001032347-2016-00396
- Event Type
- Injury
- Date Received
- August 11, 2016
- Date of Event
- July 1, 2016
- Report Date
- July 13, 2016
- Manufacturer
- BIOMET MICROFIXATION
- Product Code
- LZD
- PMA / PMN Number
- PP020016
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE PRODUCT WAS NOT RETURNED FOR AN EVALUATION, HOWEVER AN EVALUATION OF THE COMPLAINT WAS CONDUCTED. BASED ON THE EVALUATION, FIELD WAS UPDATED TO REFLECT THE OPERATIONAL PROBLEM OF: "PARTS DON'T MOVE FREELY." THIS IS REPORT TWO OF FIVE FOR THE SAME EVENT. REPORTS ONE, AND THREE THROUGH FIVE ARE REPORTED ON 1032347-2016-00395-1 AND 1032347-2016-00397-1 THROUGH 1032347-2016-00399-1.
IT WAS REPORTED THE PATIENT IS IN JAIL AND THE SURGEON BELIEVES THE EVENT WAS DUE TO THE PATIENT NOT HAVING ACCESS TO ORAL CARE. REVIEW OF DEVICE HISTORY RECORDS SHOW THE LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THE WARNINGS IN THE PACKAGE INSERT STATE THIS TYPE OF EVENT CAN OCCUR. IF ADDITIONAL INFORMATION IS OBTAINED THAT ADDS VALUE TO THE RELEVANT CONTENT OF THIS REPORT A FOLLOW-UP REPORT WILL BE SENT. REPORT TWO OF FIVE FOR THE SAME EVENT, REFERENCE 1032347-2016-00395 THROUGH 1032347-2016-00399.
IT WAS REPORTED THAT ALL PREVIOUSLY IMPLANTED PRODUCTS WERE RE-IMPLANTED INTO THE PATIENT ONCE THE PATIENT'S CONDITION WAS RESOLVED.
A REVISION SURGERY WAS REPORTED DUE TO ANKYLOSIS OVER THE JOINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 521165 | TEMPOROMANDIBULAR JOINT REPLACEMENT SYSTEM | 50MM RIGHT STANDARD MANDIBLE | LZD | BIOMET MICROFIXATION | N/A | 190880 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |