FDA Adverse Event Malfunction Summary report: N

WOLF-PAK NORMAL SALINE IV 10ML FILL IN 12ML FLUSH SYRINGE

MDR report key: 5867403 · Received August 10, 2016

Report

Report Number
1000151124-2016-00004
Event Type
Malfunction
Date Received
August 10, 2016
Date of Event
January 19, 2016
Report Date
August 9, 2016
Manufacturer
WOLF MEDICAL SUPPLY
Product Code
NGT
PMA / PMN Number
K111034
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INITIAL INFORMATION ABOUT THIS EVENT WAS PROVIDED TO AQUABILITI BY THE PRIVATE LABEL MANUFACTURER, WOLF MEDICAL SUPPLY (WMS). THE INFORMATION STATED THEIR (WMS) CUSTOMER ENCOUNTERED A SYRINGE WITH A "RUSTY" COLOR. AQUABILITI ISSUED A RETURN MATERIAL AUTHORIZATION. THE DEVICE EXHIBITING THE "RUSTY" COLOR WAS RECEIVED AND FORWARDED TO THE SYRINGE MANUFACTURER FOR EVALUATION. THE SYRINGE MANUFACTURED STATED THE REPORTED CONDITION ("RUSTY" COLOR SOLUTION) WAS OBSERVED. A METAL CONTAMINANT WAS FOUND IN THE SHOULDER OF THE SYRINGE. THE CONTAMINANT WAS SENT TO A LABORATORY FOR COMPOSITION ANALYSIS. THE RESULTS SHOWED THE CONTAMINANT WAS A MIXTURE OF IRON, CHROMIUM, ALUMINUM, MANGANESE, NICKEL, AND ZINC. THE SYRINGE MANUFACTURER RESPONDED TO US THE SOURCE OF CONTAMINANT WAS TRACED BACK TO THREE POTENTIAL DEFICIENCES AT THEIR MANUFACTURING FACILITY. CLEANING METHODS MAY BE INSUFFICIENTLY ESTABLISHED OR MAY NOT HAVE BEEN PROPERLY FOLLOWED, RAW MATERIALS USED IN THE CONSTRUCTION OF THE SYRINGE PLUNGER TIP MAY HAVE BEEN CONTAMINATED, OR POSSIBLE PRODUCTION EQUIPMENT WEAR MAY BE GENERATING METALS THAT CONTAMINATE THE MANUFACTURING PROCESS. THE SYRINGE MANUFACTURER OPENED A FORMAL CORRECTIVE ACTION FOR THIS ISSUE, AND THEIR MANUFACTURING AND QUALITY ASSURANCE PERSONNEL HAVE BEEN MADE AWARE OF THIS COMPLAINT IN AN EFFORT TO FOCUS ATTENTION ON PREVENTING THIS ISSUE.

Description of Event or Problem · 1

A CUSTOMER OF WOLF MEDICAL SUPPLY, INC. ENCOUNTERED A SYRINGE WITH A "RUSTY" COLOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
517569 WOLF-PAK NORMAL SALINE IV 10ML FILL IN 12ML FLUSH SYRINGE SALINE, VASCULAR ACCESS FLUSH NGT WOLF MEDICAL SUPPLY PS1012 KH04240

Patients

Seq Age Sex Outcome Treatment
1 Other