CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2016-09521
- Event Type
- Malfunction
- Date Received
- August 9, 2016
- Date of Event
- May 18, 2016
- Report Date
- June 20, 2016
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PHYSICIAN
Narratives
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD AND BODY TISSUE/FIBROTIC GROWTH. THE ANALYST NOTED THAT THERE WAS A SMALL AMOUNT OF TISSUE/BLOOD INSIDE THE HELIX WHICH WHEN REMOVED WITH ALCOHOL, THE HELIX OPERATED WITHIN SPECIFICATION. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE HELIX OF THE RIGHT ATRIAL LEAD COULD NOT ADVANCE. A DIFFERENT LEAD WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 510261 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-45 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR |