REVEAL LINQ
Report
- Report Number
- 9614453-2016-05074
- Event Type
- Malfunction
- Date Received
- August 9, 2016
- Date of Event
- April 4, 2016
- Report Date
- May 11, 2016
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- DSI
- PMA / PMN Number
- K132649
- Removal / Correction Number
- Z-1249-2016
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVENT SUMMARY - THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. RECOMMENDED REPLACEMENT TIME WAS TRIGGERED ON (B)(6) 2016 DUE TO IMPEDANCE. THE DAILY BATTERY TREND DATA SHOWS GRADUAL RISE OF BATTERY IMPEDANCE. PREMATURE RECOMMENDED REPLACEMENT TIME ALERT, WITHOUT CORRESPONDING BATTERY DEPLETION. THIS DEVICE WAS REPORTED AS INCLUDED IN THE FIELD ACTION. BASED ON THE INFORMATION RECEIVED PERFORMED AND WITHOUT THE RETURN OF THE PRODUCT, IT CANNOT BE CONFIRMED THAT THIS DEVICE AS DESCRIBED IN THE FIELD ACTION. IT IS INCLUDED IN THE FIELD ACTION IN THE ABUNDANCE OF CAUTION.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) DEVICE REACHED RECOMMENDED REPLACEMENT TIME (RRT) EARLIER THAN EXPECTED. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 515428 | REVEAL LINQ | DETECTOR AND ALARM, ARRHYTHMIA | DSI | IPG MFG SWITZERLAND | LNQ11 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR |