FDA Adverse Event Malfunction Summary report: N

IPC® HANDPIECE - INDIGO¿ DRILL

MDR report key: 5857689 · Received August 9, 2016

Report

Report Number
1045254-2016-00252
Event Type
Malfunction
Date Received
August 9, 2016
Date of Event
July 15, 2016
Report Date
July 15, 2016
Manufacturer
MEDTRONIC XOMED INC.
Product Code
HBE
PMA / PMN Number
K081475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT PRODUCTS: 1845020 ¿ INDIGO OTOLOGY ANGLED ATTACHMENT; MANUFACTURED DATE ¿ SEPTEMBER 24, 2015; SERIAL NUMBER - (B)(4); LOT NUMBER - 210048767; 510K NUMBER ¿ K081475. THE HANDPIECE (1845000) AND ATTACHMENT (1845020) HAVE NOT BEEN RETURNED AT THIS TIME. A PRODUCT ANALYSIS HAS NOT BEEN COMPLETED.

Additional Manufacturer Narrative · 1

DEVICE RECEIVED DATE: THE INDIGO HANDPIECE (1845000) AND ANGLED ATTACHMENT (1845020) WERE RECEIVED ON JULY 25, 2016. INDIGO HANDPIECE (1845000; SERIAL NUMBER (B)(4)): THE PRODUCT ANALYSIS INDICATES THAT THE CUSTOMER¿S COMPLAINT COULD NOT BE CONFIRMED. ONE-MINUTE PRE-REPAIR TEMPERATURE TEST RESULTS WERE AS FOLLOWS: 81 DEGREES F AT T1 AND 80 DEGREES F AT T2. THE AMBIENT TEMPERATURE WAS 76 DEGREES F. AFTER 5 MINUTES, THE TEMPERATURE AT T1 WAS 95 DEGREES F AND IT WAS 94 DEGREES F AT T2. THERE WERE NO MECHANICAL OR COSMETIC DEVIATIONS FOUND. THE HANDPIECE WAS SUCCESSFULLY TESTED TO SPECIFICATIONS. INDIGO ANGLED ATTACHMENT (1845020; SERIAL NUMBER (B)(4)): THE PRODUCT ANALYSIS INDICATES THAT THE AMBIENT TEMPERATURE WAS 75 DEGREES F. AFTER 1 MINUTE THE DEVICE WAS 96 DEGREES F AND AFTER 5 MINUTES THE DEVICE WAS 115 DEGREES F. THE BEARINGS WERE CORRODED. THE BEARINGS AND OTHER COMPONENTS WERE REPLACED. THE DEVICE WAS SUCCESSFULLY TESTED TO SPECIFICATIONS. METHOD: ACTUAL DEVICE EVALUATED (B)(4); VISUAL INSPECTION (B)(4); LABELING EVALUATION (B)(4). RESULTS: NO FAILURE DETECTED (B)(4). CONCLUSION: DEVICE REPAIRED AND RETURNED (B)(4). A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT "INTRAOPERATIVE, THE HANDPIECE AND/OR ANGLED ATTACHMENT GOT HOT IN LESS THAN 30 SECONDS. A REPLACEMENT HANDPIECE WAS USED TO COMPLETE THE CASE. THE PROCEDURE WAS NOT DELAYED AND THERE WAS NO PATIENT INJURY."