IPC® HANDPIECE - INDIGO¿ DRILL
Report
- Report Number
- 1045254-2016-00252
- Event Type
- Malfunction
- Date Received
- August 9, 2016
- Date of Event
- July 15, 2016
- Report Date
- July 15, 2016
- Manufacturer
- MEDTRONIC XOMED INC.
- Product Code
- HBE
- PMA / PMN Number
- K081475
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
CONCOMITANT PRODUCTS: 1845020 ¿ INDIGO OTOLOGY ANGLED ATTACHMENT; MANUFACTURED DATE ¿ SEPTEMBER 24, 2015; SERIAL NUMBER - (B)(4); LOT NUMBER - 210048767; 510K NUMBER ¿ K081475. THE HANDPIECE (1845000) AND ATTACHMENT (1845020) HAVE NOT BEEN RETURNED AT THIS TIME. A PRODUCT ANALYSIS HAS NOT BEEN COMPLETED.
DEVICE RECEIVED DATE: THE INDIGO HANDPIECE (1845000) AND ANGLED ATTACHMENT (1845020) WERE RECEIVED ON JULY 25, 2016. INDIGO HANDPIECE (1845000; SERIAL NUMBER (B)(4)): THE PRODUCT ANALYSIS INDICATES THAT THE CUSTOMER¿S COMPLAINT COULD NOT BE CONFIRMED. ONE-MINUTE PRE-REPAIR TEMPERATURE TEST RESULTS WERE AS FOLLOWS: 81 DEGREES F AT T1 AND 80 DEGREES F AT T2. THE AMBIENT TEMPERATURE WAS 76 DEGREES F. AFTER 5 MINUTES, THE TEMPERATURE AT T1 WAS 95 DEGREES F AND IT WAS 94 DEGREES F AT T2. THERE WERE NO MECHANICAL OR COSMETIC DEVIATIONS FOUND. THE HANDPIECE WAS SUCCESSFULLY TESTED TO SPECIFICATIONS. INDIGO ANGLED ATTACHMENT (1845020; SERIAL NUMBER (B)(4)): THE PRODUCT ANALYSIS INDICATES THAT THE AMBIENT TEMPERATURE WAS 75 DEGREES F. AFTER 1 MINUTE THE DEVICE WAS 96 DEGREES F AND AFTER 5 MINUTES THE DEVICE WAS 115 DEGREES F. THE BEARINGS WERE CORRODED. THE BEARINGS AND OTHER COMPONENTS WERE REPLACED. THE DEVICE WAS SUCCESSFULLY TESTED TO SPECIFICATIONS. METHOD: ACTUAL DEVICE EVALUATED (B)(4); VISUAL INSPECTION (B)(4); LABELING EVALUATION (B)(4). RESULTS: NO FAILURE DETECTED (B)(4). CONCLUSION: DEVICE REPAIRED AND RETURNED (B)(4). A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT "INTRAOPERATIVE, THE HANDPIECE AND/OR ANGLED ATTACHMENT GOT HOT IN LESS THAN 30 SECONDS. A REPLACEMENT HANDPIECE WAS USED TO COMPLETE THE CASE. THE PROCEDURE WAS NOT DELAYED AND THERE WAS NO PATIENT INJURY."