FRACTURE/RECONSTRUCTION SYSTEM (TRAUMAONE)
Report
- Report Number
- 0001032347-2016-00387
- Event Type
- Injury
- Date Received
- August 9, 2016
- Date of Event
- July 4, 2016
- Report Date
- July 11, 2016
- Manufacturer
- BIOMET MICROFIXATION
- Product Code
- JEY
- PMA / PMN Number
- PK081067
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- PHYSICIAN
Narratives
THE USER FACILITY IS FOREIGN; THEREFORE A FACILITY MEDWATCH REPORT WILL NOT BE AVAILABLE. REVIEW OF DEVICE HISTORY RECORDS SHOW THE LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. WITHOUT A PRODUCT RETURN, NO PRODUCT EVALUATION IS ABLE TO BE CONDUCTED. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A VALID CONCLUSION ABOUT THE CAUSE OF THIS EVENT. IF ADDITIONAL INFORMATION IS OBTAINED THAT ADDS VALUE TO THE RELEVANT CONTENT OF THIS REPORT AND/OR A CONCLUSION CAN BE DRAWN, A FOLLOW-UP REPORT WILL BE SENT. REPORT ONE OF THREE FOR THE SAME EVENT, REFERENCE 1032347-2016-00388 AND 1032347-2016-00389.
CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A VALID CONCLUSION ABOUT THE CAUSE OF THIS EVENT. A FOLLOW UP REPORT WILL BE SENT UPON COMPLETION OF THE DEVICE EVALUATION. BASED ON THE PRODUCT RETURN, THE FIELDS WERE UPDATED. THIS IS REPORT ONE OF THREE FOR THE SAME EVENT. REPORTS TWO AND THREE ARE REPORTED ON MFR #1032347-2016-00388 AND 1032347-2016-00389.
THE PRODUCT THAT WAS RETURNED IS NOT THE EXPLANTED PRODUCT/ PRODUCT IN QUESTION. IT WAS CLARIFIED THAT THE EXPLANTED PRODUCT HAS BEEN QUARANTINED AND WHAT WAS RETURNED ARE THE REMAINING SCREWS FROM THE SAME LOT THAT WERE NOT USED AT ALL. A PRODUCT EVALUATION WILL BE CONDUCTED ON THE SCREWS THAT WERE RETURNED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE PRODUCT EVALUATION. THIS IS REPORT ONE OF THREE FOR THE SAME EVENT. REPORTS TWO AND THREE ARE REPORTED ON MFR #1032347-2016-00388-2 AND 1032347-2016-00389-2.
FIVE 41-2006 SCREWS WERE RETURNED AND VISUALLY EVALUATED UNDER A DIGITAL MICROSCOPE AND IT WAS SEEN THAT THE THREADS ARE FULLY IN TACT AND THERE IS DAMAGE ON THE HEAD OF TWO SCREWS, INDICATING THAT THE SCREW MAINTAINED GOOD RETENTION WITH THE BLADE. THE SCREWS WERE FUNCTIONALLY TESTED BY USING A 2.0 DRILL (PN: 72-2018) TO DRILL THE PILOT HOLE AND THEN INSERTING THE SCREWS INTO THE WHITE OAK WOOD BLOCK. THE SCREWS FULLY SEATED INTO THE WHITE OAK WOOD BLOCK WITH EASE. THE MOST LIKELY UNDERLYING CAUSE OF THIS COMPLAINT CANNOT BE DETERMINED AS THE SCREWS FUNCTIONED AS INTENDED. SUPPLEMENTAL REPORT ONE OF THREE FOR THE SAME EVENT, REFERENCE 1032347-2016-00388-3 AND 1032347-2016-00389-3.
IT WAS REPORTED DURING AN ORTHOGNATHIC SURGERY SEVERAL SCREWS WERE UNABLE TO BE FIXATED. THE SURGEON STATED HE BELIEVES THE DRILLS MAY HAVE CREATED TOO LARGE OF A HOLE FOR THE SCREWS. THE SURGEON WAS ABLE TO SUCCESSFULLY COMPLETE THE PROCEDURE BY USING ANOTHER SYSTEM. A FORTY-FIVE MINUTE SURGICAL DELAY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 509751 | FRACTURE/RECONSTRUCTION SYSTEM (TRAUMAONE) | SCREW 2.0X6MM CROSS-DRIVE | JEY | BIOMET MICROFIXATION | N/A | 500750 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |