HYPERBARIC CHAMBER
Report
- Report Number
- 2020676-2016-00005
- Event Type
- Death
- Date Received
- August 8, 2016
- Date of Event
- August 22, 2015
- Report Date
- March 15, 2016
- Manufacturer
- SECHRIST INDUSTRIES, INC.
- Product Code
- CBF
- PMA / PMN Number
- K140559
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
SECHRIST MADE SEVERAL ATTEMPTS TO TRY AND OBTAIN THE CHAMBER FOR EVALUATION. SEVERAL PHONE CALLS WERE MADE TO DR. (B)(6) AT (B)(6) AND MESSAGES WERE NEVER RETURNED. BASED ON DEVICE HISTORY REVIEW, THIS CHAMBER WAS IN USE BEYOND THE CURRENT PVHO REGULATED CYLINDER LIFE OF 20 YEARS AND WAS MANUFACTURED IN 1989. THIS HYPERBARIC CHAMBER WAS ORIGINALLY SOLD TO (B)(6) HOSPITAL AND SECHRIST CONDUCTED ANNUAL PREVENTATIVE MAINTENANCE THROUGH 1998. AT NO TIME WAS SECHRIST MADE AWARE OF RELOCATION OF THIS CHAMBER TO ANY OTHER FACILITY. (B)(6) FIRE INVESTIGATION STATED THAT DR. (B)(6) PURCHASED THE CHAMBER ON (B)(6). BASED ON (B)(6) FIRE INVESTIGATIONS SECHRIST INDUSTRIES ALSO LEARNED THAT DR. (B)(6) PLACED A BATTERY OPERATED FAN IN THE CHAMBER WHILE IN USE. INTRODUCTION OF ANY TYPE OF HEAT SOURCE INTO A HYPERBARIC CHAMBER IS STRICTLY AGAINST MANUFACTURER RECOMMENDATIONS AS WELL AS NFPA99 HEALTH CARE FACILITIES CODE, CHAPTER 14 HYPERBARIC FACILITIES. : DEVICE NOT PROVIDED FOR EVALUATION.
SECHRIST INDUSTRIES RECEIVED AN EMAIL FROM (B)(6) WITH (B)(6) NEWS STATING THAT THEY WOULD BE AIRING A STORY ABOUT A FIRE THAT TOOK PLACE INSIDE A HYPERBARIC CHAMBER AT (B)(6). REPORTEDLY, (B)(6) BROUGHT HER DOG TO THE VETERINARIAN CLINIC WHERE DR. (B)(6) GAVE THE DOG OXYGEN THERAPY BY PUTTING THE DOG INSIDE A HYPERBARIC CHAMBER. DURING THERAPY THE CHAMBER CAUGHT ON FIRE, KILLING THE DOG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 505696 | HYPERBARIC CHAMBER | HYPERBARIC CHAMBER | CBF | SECHRIST INDUSTRIES, INC. | 2500B/BR |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Death |