FDA Adverse Event Other Summary report: N

MULTIVIEW WORKSTATION WITH CRS OPTION

MDR report key: 585431 · Received March 21, 2005

Report

Report Number
1220063-2005-00003
Event Type
Other
Date Received
March 21, 2005
Date of Event
February 12, 2003
Report Date
March 1, 2005
Manufacturer
DRAEGER MEDICAL SYSTEMS, INC.
Product Code
DRT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

TWO DIFFERENT PATIENTS GOT THEIR ECG STORED ON THE LAST PATIENT'S IDENTITY IN THE MULTIVIEW WORKSTATION WITH CRS (CARDIAC REVIEW STATION) OPTION. THE FIRST PATIENT'S ECG SHOWS CHANGES, WHICH SHOULD LEAD TO AN ANGIOGRAPHY IF IT HAD BEEN THE LAST PATIENTS ECG. THIS WAS DISCOVERED BEFORE ANY PREVENTIVE ACTIONS. THE CUSTOMER'S CERTIFIED BIOMEDICAL ENGINEER'S FURTHER COMMENTS: THE REASON FOR THIS FAULT IS THAT FIRST PATIENT WASN'T DISCHARGED FROM THE SYSTEM BEFORE LAST WAS CONNECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MULTIVIEW WORKSTATION WITH CRS OPTION CENTRAL STATION PATIENT MONITOR DRT DRAEGER MEDICAL SYSTEMS, INC. UNK NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN