FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
MDR report key: 5853610
·
Received August 8, 2016
Report
- Report Number
- 3015876-2016-00927
- Event Type
- Malfunction
- Date Received
- August 8, 2016
- Date of Event
- July 15, 2016
- Report Date
- August 8, 2016
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K142430
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PMA 510(K): OF THE INITIAL MEDWATCH REPORT INDICATED: 510K - K130456. THE INITIAL MEDWATCH REPORT SHOULD HAVE INDICATED: 510K - K142430
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND WAS UNABLE TO DUPLICATE THE REPORTED ISSUE. PHYSIO REPLACED THE BATTERY CONNECTOR PINS AS A PRECAUTION AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE. THE CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THEIR DEVICE WAS LOSING POWER ON A CONSISTENT BASIS. THERE WAS NO REPORT OF ANY PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 506202 | LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR | DEFIBRILLATORS, AUTOMATIC, EXTERNAL | MKJ | PHYSIO-CONTROL, INC | 15 |