FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR

MDR report key: 5853610 · Received August 8, 2016

Report

Report Number
3015876-2016-00927
Event Type
Malfunction
Date Received
August 8, 2016
Date of Event
July 15, 2016
Report Date
August 8, 2016
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K142430
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PMA 510(K): OF THE INITIAL MEDWATCH REPORT INDICATED: 510K - K130456. THE INITIAL MEDWATCH REPORT SHOULD HAVE INDICATED: 510K - K142430

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND WAS UNABLE TO DUPLICATE THE REPORTED ISSUE.  PHYSIO REPLACED THE BATTERY CONNECTOR PINS AS A PRECAUTION AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING.  THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE. THE CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEIR DEVICE WAS LOSING POWER ON A CONSISTENT BASIS. THERE WAS NO REPORT OF ANY PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
506202 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR DEFIBRILLATORS, AUTOMATIC, EXTERNAL MKJ PHYSIO-CONTROL, INC 15