FDA Adverse Event Malfunction Summary report: N

VITEK® 2 ANC TEST KIT

MDR report key: 5851691 · Received August 5, 2016

Report

Report Number
1950204-2016-00085
Event Type
Malfunction
Date Received
August 5, 2016
Report Date
July 4, 2016
Manufacturer
BIOMERIEUX, INC
Product Code
JSP
PMA / PMN Number
K910666
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

AN INVESTIGATION INTO A PRODUCT ISSUE WHILE USING THE VITEK® ANC TEST KIT WAS PERFORMED. THE CUSTOMER REPORTED DISCREPANT RESULTS WITH A QC SAMPLE. TESTING WAS PERFORMED ON 4 ANC CARDS (CUSTOMER LOT TWICE: 244379420 AND RANDOM LOT TWICE: 244366910 ON CBA IN AN ANAEROBIC ATMOSPHERE) ON EACH STRAIN TESTED, BOTH THE CUSTOMER AND AN INTERNAL REFERENCE STRAIN. THE CUSTOMER STRAIN WAS TESTED. THE DISCREPANT RESULTS WERE DUPLICATED. WHEN THE INTERNAL REFERENCE STRAIN WAS TESTED, THE RESULTS WERE AS EXPECTED. QC WAS CONFORMING IN-HOUSE WITH ALL LOTS TESTED WHATEVER THE MEDIA USED FOR THE INTERNAL REFERENCE STRAIN. THEREFORE, THE MOST LIKELY CAUSE OF THIS EVENT IS A DEGRADED STRAIN FROM THE CUSTOMER. ANC CARDS PERFORMED AS INTENDED AND NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

A CUSTOMER NOTIFIED BIOMERIEUX OF A MIS-IDENTIFICATION WHEN USING THE VITEK 2 ANC TEST KIT. AN ATCC QUALITY CONTROL SAMPLE OF CLOSTRIDIUM SEPTICUM (ATCC 12464) WAS RUN AND GAVE THE RESULTS OF DGLU+, BMAN+, MTE+, AND OPS-. THE EXPECTED RESULTS ARE DGLU-, BMAN-, MTE-, AND OPS+. WHEN SPECIFICALLY ASKED THE CUSTOMER NOTED NO INJURY OR DEATH HAPPENED AS A RESULT OF THIS INCIDENT. AN INVESTIGATION HAS BEEN INITIATED BY BIOMERIEUX TO INVESTIGATE THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
502989 VITEK® 2 ANC TEST KIT VITEK® 2 ANC TEST KIT JSP BIOMERIEUX, INC 244379420

Patients

Seq Age Sex Outcome Treatment
1