OT ULTRA2 METER
Report
- Report Number
- 3008382007-2016-39994
- Event Type
- Injury
- Date Received
- August 4, 2016
- Report Date
- July 28, 2016
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K053529
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PATIENT
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
THE LAY USER/PATIENT¿S METER AND TEST STRIPS HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE METER FAILED TESTING. THE REPORTED ISSUE WAS CONFIRMED. THE TEST STRIPS PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2016, THE LAY-USER/PATIENT CONTACTED LIFESCAN (LFS) USA, ALLEGING THAT HER ONETOUCH ULTRA2 METER WAS READING INACCURATELY ERRATIC. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER SERVICE REPRESENTATIVE (CSR) DOCUMENTATION. THE PATIENT REPORTED THAT THE ALLEGED METER INACCURACY BEGAN ON (B)(6) 2016 AT 4:08PM. THE PATIENT REPORTED OBTAINING BLOOD GLUCOSE READINGS OF "316, 312, 162, 311 AND 138 MG/DL" WITH THE SUBJECT METER, PERFORMED WITHIN 20 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN'S CRITERIA FOR PRECISION. THE PATIENT INFORMED THE CSR THAT SHE MANAGES HER DIABETES WITH A FIXED DOSE OF INSULIN (TYPE/DOSE NOT SPECIFIED) AND CLAIMED SHE TOOK EXERCISE ON (B)(6) 2016 AT 6:00PM IN RESPONSE TO THE ALLEGED ISSUE. THE PATIENT REPORTED THAT 19 DAYS AFTER THE ALLEGED ISSUE BEGAN SHE DEVELOPED SYMPTOMS OF "SHAKINESS AND SWEATS" BUT DENIED RECEIVING ANY MEDICAL TREATMENT. DURING TROUBLESHOOTING, THE CSR CONFIRMED THAT THE UNIT OF MEASURE WAS SET CORRECTLY ON THE SUBJECT METER. THE CSR NOTED THAT THE SAME APPROVED SAMPLE SITE WAS USED FOR TESTING AND THAT THE CORRECT TESTING PROCESS WAS BEING FOLLOWED. THE CSR NOTED THAT THE PATIENT DID NOT HAVE CONTROL SOLUTION AVAILABLE TO TEST THE SUBJECT METER. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT REPORTEDLY DEVELOPED SYMPTOMS SUGGESTIVE OF A SERIOUS INJURY AFTER OBTAINING THE ALLEGED INACCURATE RESULTS WITH THE SUBJECT METER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 499139 | OT ULTRA2 METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 4008693 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Life Threatening |