O-ARM 1000 IMAGING SYSTEM

MDR report key: 5845257 · Received August 3, 2016

Report

Report Number
1723170-2016-01691
Event Type
Malfunction
Date Received
August 3, 2016
Date of Event
August 27, 2014
Report Date
August 3, 2016
Manufacturer
MEDTRONIC NAVIGATION, INC. (LITTLETON)
Product Code
OXO
PMA / PMN Number
K050996
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A MEDTRONIC REPRESENTATIVE VISITED THE SITE TO EXAMINE THE MEDTRONIC IMAGING SYSTEM, RESULTING IN THE DECISION TO RETRIEVE LOG FILES AND RETURN THEM TO THE MANUFACTURER FOR ANALYSIS SINCE THE ORIGINAL ISSUE COULD NOT BE REPLICATED IN SITU. AFTER OFFLOADING THE FILES, THE REPRESENTATIVE PERFORMED AN IMAGING SYSTEM CHECK-OUT AND TESTED AREAS PASSED. SYSTEM PERFORMED AS INTENDED. THE SOFTWARE INVESTIGATION FOUND THAT THE REPORTED EVENT WAS RELATED TO A SOFTWARE ISSUE. THIS ISSUE WAS DOCUMENTED IN A MEDTRONIC NAVIGATION SOFTWARE ANOMALY TRACKING DATABASE. THIS EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW AS DISCUSSED WITH THE FDA (B)(4) DISTRICT OFFICE ON APRIL 7, 2016 VIA MEDTRONIC NAVIGATION, INC. RESPONSE TO 3/17/2016 (B)(4). THIS REVIEW WAS PERFORMED AS A RESULT OF RECENT CHANGES/IMPROVEMENTS TO THE MEDTRONIC NAVIGATION, INC. MEDICAL DEVICE REPORT (MDR) REVIEW PROCESS AND IS FOR NON-ADVERSE EVENT REPORTABLE MALFUNCTIONS WITHIN THE LAST TWO YEARS WHERE THE USE OF THE O-ARM IMAGING SYSTEM WAS ABORTED. THIS PROCESS CHANGE RESULTED IN AN OVERALL INCREASE IN THE NUMBER OF MDRS FILED BY MEDTRONIC NAVIGATION, INC., SINCE APRIL 2016. IT SHOULD BE NOTED THAT THIS INCREASE IS NOT THE RESULT OF THE DISCOVERY OF NEW PRODUCT PROBLEMS, BUT RATHER THE RESULT OF A BROADER INTERPRETATION AND APPLICATION OF THE REGULATIONS WITHIN OUR PROCESSES.

Description of Event or Problem · 1

A SITE REPRESENTATIVE REPORTED THAT, WHILE IN A SPINAL FUSION PROCEDURE, THE MEDTRONIC IMAGING SYSTEM WAS UNABLE TO TRANSFER IMAGES TO THE MEDTRONIC NAVIGATION SYSTEM. SWAPPING THE NETWORK CABLE DID NOT RESOLVE THE ISSUE. THE SURGEON DECIDED TO ABORT THE USE OF THE MEDTRONIC IMAGING AND NAVIGATION SYSTEM. THERE WAS A REPORTED DELAY TO THE PROCEDURE OF LESS THAN 1 HOUR DUE TO THIS ISSUE. THERE WAS NO IMPACT ON PATIENT OUTCOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
495439 O-ARM 1000 IMAGING SYSTEM IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE OXO MEDTRONIC NAVIGATION, INC. (LITTLETON) BI-700-00027-120

Patients

Seq Age Sex Outcome Treatment
1 55 YR