LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2016-05204
- Event Type
- Injury
- Date Received
- August 3, 2016
- Date of Event
- July 2, 2016
- Report Date
- August 1, 2016
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
THERE WAS NO DEATH OR DEVICE MALFUNCTION ASSOCIATED WITH THE INAPPROPRIATE DEFIBRILLATION EVENT. DEVICE EVALUATION SUMMARY: MONITOR SN (B)(4) AND ELECTRODE BELT SN (B)(4) WERE RETURNED AND EVALUATED AT THE DISTRIBUTOR, IN ACCORDANCE WITH PROCEDURES RECOMMENDED BY ZOLL MANUFACTURING CORPORATION. THE EVALUATION INCLUDED REVIEW OF DOWNLOADED SOFTWARE FLAG FILES (ATTACHED) ON THE DAY OF THE EVENT AND INCOMING FUNCTIONAL TESTING. THE REVIEW OF THE SOFTWARE FLAGS CONSISTED OF AN ANALYSIS OF THE DOWNLOADED DATA TO IDENTIFY ANY FAULT FLAGS OR UNUSUAL PATTERNS OF SOFTWARE FLAGS. DOWNLOAD FAULT FLAGS WERE PRESENT, BUT DOWNLOADING IS A SECONDARY FUNCTION OF THE DEVICE AND DOES NOT AFFECT THE DEVICE'S ABILITY TO DETECT AND TREAT A PATIENT. THE SOFTWARE FLAG FILES DID NOT SUGGEST A DEVICE MALFUNCTION THAT WOULD CONTRIBUTE TO THE INAPPROPRIATE TREATMENT. DURING THE INCOMING FUNCTIONAL TESTING, A 1HZ SIMULATED NORMAL SINUS RHYTHM SIGNAL WAS APPLIED TO THE ECG ELECTRODES, FOLLOWED BY A 5HZ SIMULATED TREATABLE ARRHYTHMIA SIGNAL WHICH VERIFIED PROPER PERFORMANCE OF THE DETECTION ALGORITHM. DURING THE TRANSITION TO THE 5HZ SIGNAL, THE DEVICE WAS CONFIRMED TO PROPERLY ENTER INTO A TREATMENT SEQUENCE WHICH INCLUDES A VERIFICATION OF THE TACTILE VIBRATION ALARM, AUDIO MESSAGING, AND SIREN ALARMS, AS WELL AS A TEST OF THE PULSE DELIVERY CIRCUITRY. THE PULSE DELIVERY CIRCUITRY TEST VERIFIED PROPER CHARGING OF THE HIGH VOLTAGE CAPACITORS AND PROPER DELIVERY OF FIVE FULL ENERGY 150J BIPHASIC PULSES. THE FUNCTIONAL TESTING CONFIRMED PROPER RESPONSE BUTTON FUNCTIONALITY, ECG ACQUISITION, DETECTION ALGORITHM PERFORMANCE, AND PULSE DELIVERY FUNCTIONALITY. DEVICE MANUFACTURE DATE: MONITOR: SN (B)(4): 07/31/2015 REUSE, ELECTRODE BELT: SN (B)(4): 04/14/3014 INITIAL USE. THE INVESTIGATION INTO THE EVENT CONCLUDES THAT THERE WAS NO DEVICE MALFUNCTION. A CAUSE AND EFFECT ANALYSIS WAS CONDUCTED USING ALL OF THE AVAILABLE INFORMATION WHICH INCLUDES THE INCIDENT REPORT, DEVICE EVALUATION, SOFTWARE FLAG FILES, AND ECG STRIPS. THE PRIMARY CAUSE OF THE INAPPROPRIATE SHOCK WAS INAPPROPRIATE BUTTON USE BY THE PATIENT DURING THE FALSE DETECTION. THE ECG ANALYSIS, CONDUCTED BY TRAINED ECG TECHNICIANS, IDENTIFIED THE PRIMARY CAUSE OF THE FALSE DETECTION WAS RAPID ATRIAL FIBRILLATION. THE RAPID RATE SATISFIED THE RATE DETECTOR OF THE DETECTION ALGORITHM. INAPPROPRIATE DEFIBRILLATIONS ARE AN ANTICIPATED RISK ASSOCIATED WITH THE USE OF THE LIFEVEST. PATIENTS ARE INSTRUCTED THROUGH ALARMS, VOICE MESSAGES, IFU, AND TRAINING TO PRESS THE RESPONSE BUTTONS TO PREVENT AN INAPPROPRIATE DEFIBRILLATION. THE CURRENT COMMERCIAL INAPPROPRIATE DEFIBRILLATION RATE IS CONSISTENT WITH THE OBSERVED RATE DURING THE PIVOTAL CLINICAL TRIAL G960083 (0.69% PER PATIENT-MONTH WITH 90% CONFIDENCE). A SUMMARY OF THE SAFETY AND EFFECTIVENESS DATA (SSED), INCLUDING THE INAPPROPRIATE DEFIBRILLATION SAFETY OBJECTIVE SUPPORTING FDA'S APPROVAL OF THE LIFEVEST, CAN BE FOUND AT HTTP://WWW.ACCESSDATA.FDA.GOV/CDRH_DOCS/PDF/P010030B.PDF. THE LIFEVEST DETECTION ALGORITHM COMPLIES WITH IEC 60601-2-4 PERFORMANCE REQUIREMENTS FOR SENSITIVITY AND SPECIFICITY.
A US DISTRIBUTOR CONTACTED ZOLL TO REPORT THAT A PATIENT EXPERIENCED AN INAPPROPRIATE DEFIBRILLATION EVENT CONSISTING OF ONE SHOCK. THE RESPONSE BUTTONS WERE PRESSED AFTER THE TREATMENT WAS DELIVERED. RAPID ATRIAL FIBRILLATION CONTRIBUTED TO THE FALSE DETECTION. THE RAPID RATE SATISFIED THE RATE DETECTOR OF THE DETECTION ALGORITHM. THE PATIENT DID NOT SEEK MEDICAL ATTENTION FOLLOWING THE EVENT. THE PATIENT CONTINUES WEARING THE LIFEVEST. THERE WAS NO DEATH OR DEVICE MALFUNCTION ASSOCIATED WITH THE INAPPROPRIATE DEFIBRILLATION EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 494933 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Other |