FDA Adverse Event Malfunction Summary report: N

PELORIS RAPID TISSUE PROCESSOR

MDR report key: 5842287 · Received August 3, 2016

Report

Report Number
8020030-2016-00037
Event Type
Malfunction
Date Received
August 3, 2016
Date of Event
June 30, 2016
Report Date
July 4, 2016
Manufacturer
LEICA BIOSYSTEMS MELBOURNE PTY. LTD.
Product Code
IEO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ON-SITE ASSESSMENT OF THE OPERATION AND FUNCTION OF THE INSTRUMENT WAS CONDUCTED BY A LEICA FIELD SERVICE ENGINEER (FSE) ON 06 JULY 2016. THE FSE PERFORMED SYSTEM VERIFICATION TESTS, FOR WHICH ALL RESULTS WERE SATISFACTORY; AND THE INSTRUMENT WAS FOUND TO BE OPERATING WITHIN SPECIFICATION. THE FSE ADVISED THE COMPLAINANT TO REPLACE ALL THE REAGENTS ON THE INSTRUMENT, INCLUDING THE WAXES, PRIOR TO PROCESSING OF DIAGNOSTIC TISSUE SAMPLES. MANUFACTURER EVALUATION OF THE AVAILABLE INSTRUMENT LOGS SHOWED THAT: AT 08:58AM ON (B)(6) 2016, A USER COMMENCED A MANUAL FILL OPERATION OF RETORT A USING THE REAGENT FROM BOTTLE 6 (ETHANOL). THIS OPERATION COMPLETED SUCCESSFULLY AT 08:59AM ON (B)(6) 2016. AT 09:04AM ON (B)(6) 2016, A USER COMMENCED A MANUAL DRAIN OPERATION OF RETORT A TO BOTTLE 6 (ETHANOL). THIS OPERATION COMPLETED SUCCESSFULLY AT 09:06AM ON (B)(6) 2016. A CODE RECORDED ON TWO (2) OCCASIONS AT 09:07AM ON (B)(6) 2016 INDICATES THAT THE "A PROGRAM" PROTOCOL, WHICH WAS BEING SCHEDULED BY THE USER IN RETORT A, COULD NOT BE RUN BECAUSE THE FIRST REAGENT SCHEDULED FOR THIS PROTOCOL (FORMALIN) WAS INCOMPATIBLE WITH RESIDUE REMAINING IN THE RETORT FROM THE MOST RECENTLY USED REAGENT, WHICH WAS ETHANOL IN THIS INSTANCE. THE FOLLOWING INFORMATION WAS DISPLAYED IN ASSOCIATION WITH THIS ERROR CODE: "INCOMPATIBLE REAGENT IN RETORT. CLEAN RETORT OR EDIT PROTOCOL, RETORT A." AT 09:08AM ON (B)(6) 2016, A USER EDITED THE "A PROGRAM" PROTOCOL TO COMMENCE AT THE FINAL DEHYDRATION STEP. THIS MODIFIED PROTOCOL COMPRISED ONE (1) DEHYDRATION STEP, THREE (3) CLEARING STEPS AND THREE (3) WAX INFILTRATION STEPS. BASED ON INFORMATION PROVIDED BY THE COMPLAINANT, IT WAS DETERMINED THAT THE SUB-OPTIMAL TISSUE PROCESSING REPORTED WAS DERIVED FROM THE MODIFIED "A PROGRAM" PROTOCOL STARTED IN RETORT A AT 09:09AM ON (B)(6) 2016, WHICH COMPLETED SUCCESSFULLY AT 13:52AM ON (B)(6) 2016. THIS PROTOCOL COMPRISED 300 CASSETTES BASED ON INFORMATION ENTERED BY THE USER INTO THE INSTRUMENT SOFTWARE. INVESTIGATION OF THIS COMPLAINT FOUND THAT THE INSTRUMENT FUNCTIONED AS DESIGNED DURING EXECUTION OF THE MODIFIED "A PROGRAM" PROTOCOL STARTED IN RETORT A AT 09:09AM ON (B)(6) 2016. THE ROOT CAUSE OF THE SUB-OPTIMAL TISSUE PROCESSING REPORTED WAS A USE ERROR, WHICH OCCURRED WHEN THE "A PROGRAM" PROTOCOL WAS MODIFIED TO COMPRISE A SINGLE DEHYDRATION STEP ONLY PRIOR TO THREE (3) CLEARING STEPS AND THREE (3) WAX INFILTRATION STEPS. A SINGLE DEHYDRATION STEP IS NOT SUFFICIENT TO DEHYDRATE THE TISSUE SAMPLES. INCOMPLETE DEHYDRATION OF THE TISSUE SAMPLES WOULD HAVE LEAD TO INCOMPLETE CLEARING AND WAX INFILTRATION OF THE TISSUE SAMPLES AND RESULTED IN THE SUB-OPTIMAL TISSUE PROCESSING REPORTED.

Description of Event or Problem · 1

LEICA BIOSYSTEMS RECEIVED A COMPLAINT REGARDING INEFFECTIVE DEHYDRATION OF TISSUE SAMPLES. ON 25 JULY 2016, LEICA BIOSYSTEMS MELBOURNE RECEIVED CONFIRMATION FROM CUSTOMER CARE THAT ALL TISSUE SAMPLES INVOLVED IN THIS EVENT WERE ABLE TO BE DIAGNOSED WITHOUT FURTHER MEDICAL OR SURGICAL INTERVENTION. INVESTIGATION OF THIS COMPLAINT BY LEICA BIOSYSTEMS IS IN PROGRESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
496991 PELORIS RAPID TISSUE PROCESSOR AUTOMATED TISSUE PROCESSOR IEO LEICA BIOSYSTEMS MELBOURNE PTY. LTD. PELORIS II

Patients

Seq Age Sex Outcome Treatment
1