BD LOGIC BLOOD GLUCOSE MONITOR

MDR report key: 584190 · Received March 15, 2005

Report

Report Number
2243072-2005-00084
Event Type
Other
Date Received
March 15, 2005
Date of Event
February 16, 2005
Report Date
March 15, 2005
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

CALLED TO REPORT COMPARING OLD METER WITH THE NEW ONE AND GETTING VERY DIFFERENT RESULTS. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVG. READING (263) AS REF. POINT VS. BD READINGS OF 171, 308, 312 YIELDING DATA POINTS IN THE D ZONE OF THE GRID, OUTSIDE OF ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD LOGIC BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 4064274

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other