FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 5840415 · Received August 2, 2016

Report

Report Number
2032227-2016-14134
Event Type
Malfunction
Date Received
August 2, 2016
Date of Event
July 18, 2013
Report Date
July 18, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD COMPROMISED FORCE SENSOR SYSTEM ALARM DURING PRIME TEST AND MOTOR ERROR ALARM DURING THE BASIC OCCLUSION TEST DUE TO FAULTY FORCE SENSOR RESISTOR. MOTOR PASSED MOTOR TEST. THE INSULIN PUMP WAS UNABLE TO PERFORM THE NO DELIVERY TEST DUE TO COMPROMISED FORCE SENSOR SYSTEM ALARM. DRIVE SUPPORT DISK WAS INSPECTED AND NO ANOMALY WAS NOTED DURING TESTING. NOTE: THIS IS A REMEDIATION MDR. MEDTRONIC DIABETES IMPLEMENTED REVISED MDR REPORTABILITY CRITERIA EFFECTIVE ON JULY 1, 2014. SUBSEQUENTLY, MEDTRONIC DIABETES CONDUCTED A ONE YEAR RETROSPECTIVE REVIEW OF COMPLAINTS. THIS EVENT WAS RETROSPECTIVELY IDENTIFIED TO BE REPORTABLE BASED ON THE REVISED MDR REPORTABILITY CRITERIA.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY RECEIVED COMPROMISED FORCE SENSOR SYSTEM ALARM AND MOTOR ERROR ALARM. THE CUSTOMER'S BLOOD GLUCOSE WAS 264 MG/DL AT THE TIME OF INCIDENT. THE CUSTOMER STATED THAT THE INSULIN PUMP WAS NOT DROPPED OR BUMPED PRIOR TO THE COMPROMISED FORCE SENSOR SYSTEM ALARM. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP APPEARED NORMAL. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WILL NEED TO BE REPLACED. THE CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO BACK-UP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
493779 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LNAS

Patients

Seq Age Sex Outcome Treatment
1 30 YR