FDA Adverse Event Malfunction Summary report: N

HLM TUBING SET W/BIOLINE COATING

MDR report key: 5840271 · Received August 2, 2016

Report

Report Number
8010762-2016-00496
Event Type
Malfunction
Date Received
August 2, 2016
Report Date
July 11, 2016
Manufacturer
MAQUET CARDIOPULMONARY AG
Product Code
DTZ
PMA / PMN Number
K132829
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

MAQUET MEDICAL SYSTEMS, USA SUBMITS THIS REPORT ON BEHALF OF THE LEGAL MANUFACTURER OF THE DEVICE MAQUET (B)(4). THE DEVICE HAS BEEN REQUESTED BUT NOT YET RECEIVED. A FOLLOW-UP MEDWATCH WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. ADDITIONAL INFORMATION: THE PRODUCT MENTIONED UNDER SECTION IS A TUBING SET AND THE INCLUDED AFFECTED COMPONENT HAS THE CONTRIBUTING DESIGN FUNCTION OF THE QUADROX-ID SMALL ADULT WHICH IS REGISTERED UNDER 510(K): K132829.

Additional Manufacturer Narrative · 1

THE PRODUCT WAS NOT AVAILABLE TO RETURN; THEREFORE A MANUFACTURER LABORATORY INVESTIGATION WAS NOT POSSIBLE. THE INCIDENT REPORT SAYS THAT 3 OXYGENATORS WERE EXCHANGED FOR THE SAME PATIENT. ONE OF THE EXCHANGED OXYGENATORS WAS RETURNED UNDER RECORD ID (B)(4) AND WAS INVESTIGATED AS FOLLOWS. THE PRODUCT WAS TESTED FOR ITS O2, CO2 TRANSFER RATE AND PRESSURE DROP BEHAVIOR AT MAXIMUM FLOW AND PASSED ALL TEST STEPS SUCCESSFULLY. NO ABNORMALITIES WERE DETECTED. BASED ON THIS THE REPORTED FAILURE "PAO2 DROPS" COULD NOT BE CONFIRMED. THE CAUSE OF THIS FAILURE WAS DETERMINED TO NOT BE ATTRIBUTED TO A DEVICE RELATED MALFUNCTION. THE MOST PROBABLE CAUSE OF THE INCIDENT IS UNKNOWN. BASED ON THE ABOVE MENTIONED INVESTIGATION RESULTS OF A DEVICE WITH CONTRIBUTING DESIGN FUNCTION THE FAILURE REPORTED WITHIN THIS INCIDENT REPORT COULD NOT BE CONFIRMED. SINCE THE REPORTED FAILURE DID NOT CONTRIBUTE TO A DEATH OR SERIOUS INJURY NO CORRECTIVE ACTION IS NEEDED. IN ADDITION AT THIS TIME IT CANNOT BE CONCLUDED THAT THIS IS A SYSTEMIC ERROR. NO CORRECTIVE ACTION IS NEEDED. THE DATA IS ALSO BEING HANDLED THROUGH A DESIGNATED MAQUET CARDIOPULMONARY TRENDING AND APPLICABLE INVESTIGATION PROCESS. DUE TO THIS NO FURTHER INVESTIGATION INITIATIONS WILL BE COMPLETED AT THIS TIME.

Description of Event or Problem · 1

DESCRIPTION FROM THE CUSTOMER REPORT: "PATIENT PAO2 DROPS AFTER ABOUT 7 DAYS OF USE. THIS HAS BEEN THE CASE ON THE SAME PATIENT AND THE CLINICIANS ARE CHANGING TO THEIR 2ND CIRCUIT. (B)(4) IS FIRST CIRCUIT. (COMPLAINT # (B)(4)) (B)(4) IS SECOND CIRCUIT (COMPLAINT # (B)(4)) (B)(4) (OXY LOT # 70108712) IS 3RD CIRCUIT (AND IS THE ONE BEING RETURNED ; COMPLAINT # (B)(4)) 4TH CIRCUIT HAS NO REPORTED ISSUES. ADDITIONAL INFO: PATIENT HAS AUTOIMMUNE ANEMIA; PROPOFOL IS OCCASIONALLY USED FOR ANESTHESIA; OXY UDI'S ARE NOT RECORDED, ONLY LOT NUMBERS BY THE CUSTOMER; NO PATIENT CONSEQUENCES WERE DESCRIBED". (B)(4).

Description of Event or Problem · 1

(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
494442 HLM TUBING SET W/BIOLINE COATING OXYGENATOR, CARDIOPULMONARY BYPASS DTZ MAQUET CARDIOPULMONARY AG BEQ-HLS 5050 USA 70110362

Patients

Seq Age Sex Outcome Treatment
1