FDA Adverse Event Injury Summary report: N

HAKIM PROGRAMMABLE MICRO VALVE

MDR report key: 5839894 · Received August 2, 2016

Report

Report Number
1226348-2016-10542
Event Type
Injury
Date Received
August 2, 2016
Date of Event
March 2, 2016
Manufacturer
CODMAN & SHURTLEFF, INC. / MEDOS S.A.
Product Code
JXG
PMA / PMN Number
K980778
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UPON COMPLETION OF THE INVESTIGATION A FOLLOW UP REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

(B)(4). UPON COMPLETION OF THE INVESTIGATION IT WAS NOTED THAT THE IMAGES WERE TAKEN OF THE ¿AS RECEIVED¿ VALVE. THE POSITION OF THE CAM WHEN VALVE WAS RECEIVED WAS 70MMH2O. THE VALVE WAS HYDRATED FOR 24 HOURS. THE VALVE WAS VISUALLY INSPECTED: NO DEFECTS WERE NOTED. THE VALVE WAS TESTED FOR PROGRAMMING. WITH PROGRAMMER 82-3126 WITH SERIAL NUMBER (B)(4), THE VALVE PASSED THE TEST. THE VALVE WAS FLUSHED, THE VALVE PASSED THE TEST NO OCCLUSION WAS NOTED. THE VALVE WAS LEAK TESTED, NO LEAKS WERE NOTED. THE VALVE WAS REFLUX TESTED. THE VALVE PASSED THE TEST. THE CATHETER WAS IRRIGATED, NO OCCLUSION WAS NOTED. THE VALVE WAS DRIED. THE VALVE WAS THEN PRESSURE TESTED, THE VALVE PASSED THE TEST. REVIEW OF THE HISTORY DEVICE RECORDS CONFIRMED THE VALVE PRODUCT CODE 82-3112 WITH LOT CRKCWB, CONFORMED TO THE SPECIFICATIONS WHEN RELEASED TO STOCK IN 29TH SEPTEMBER 2014. NO ROOT CAUSE COULD BE DETERMINED, AS THE PROBLEM REPORTED BY THE CUSTOMER COULD NOT BE DUPLICATED. BASED ON THE RESULTS OF THIS INVESTIGATION NO FURTHER ACTION IS REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME THIS COMPLAINT IS CLOSED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE VALVE SHOWED STATOR DISLODGEMENT. THE VALVE WAS EXPLANCTED AND REPLACED WITH A LIKE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
492385 HAKIM PROGRAMMABLE MICRO VALVE SHUNT, CENTRAL NERVOUS SYSTEM & COMPS JXG CODMAN & SHURTLEFF, INC. / MEDOS S.A. CRKCWB

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention