FDA Adverse Event Malfunction Summary report: N

ENDOSONICS AUTOMATIC PULL BACK DEVICE

MDR report key: 5835108 · Received July 29, 2016

Report

Report Number
2939520-2016-00051
Event Type
Malfunction
Date Received
July 29, 2016
Date of Event
March 8, 2016
Report Date
March 8, 2016
Manufacturer
ENDOSONICS CORP.
Product Code
ITX
PMA / PMN Number
K962293
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS CASE WAS REVIEWED AND INVESTIGATED ACCORDING TO THE MANUFACTURER'S POLICY. NO PATIENT INVOLVEMENT; THE ISSUE WAS FOUND DURING A MAINTENANCE CHECK IN THE MANUFACTURER'S WAREHOUSE. THIS IS A REFURBISHED PRODUCT. THE ORIGINAL SERIAL # WAS (B)(4). THERE ARE NO LOT NUMBERS ASSOCIATED WITH THESE DEVICES. THERE IS NO EXPIRATION DATE ASSOCIATED WITH THIS DEVICE. THE FIELD SERVICE ENGINEER TESTED THE DEVICE. NO FACILITY INVOLVEMENT. THIS IS AN EMPLOYEE OF THE MANUFACTURER. THIS IS A REFURBISHED PRODUCT AND DATE NOTED IS DATE IT WAS REFURBISHED. THE ORIGINAL DEVICE MANUFACTURE DATE WAS 03/10/2005. THE DEVICE WILL BE TRANSFERRED TO THE REFURBISH DEPARTMENT TO EVALUATE FOR POSSIBLE REPAIR AND RE-CERTIFICATION. THIS COMPLAINT WILL BE MONITORED AS PART OF COMPLAINT DATA ANALYSIS.

Description of Event or Problem · 1

DURING A ROUTINE INSPECTION THE MANUFACTURER'S FIELD SERVICE ENGINEER FOUND THE DEVICE'S PULLBACK SPEED APPEARED FASTER THAN THE STANDARD SPEED SETTING. THE DEVICE WAS REQUESTED TO BE RETURNED FOR ANALYSIS. THE RETURNED DEVICE WAS INVESTIGATED ACCORDING TO THE MANUFACTURER'S POLICY. VISUAL INSPECTION DISCOVERED A MISSING BATTERY COVER AND THE BATTERY ITSELF. NO FURTHER DAMAGE WAS OBSERVED ON THE DEVICE. A KNOWN GOOD 9V BATTERY (MULTIMETER JMD # (B)(4)) WAS INSTALLED INTO THE DEVICE. THE DEVICE WAS FUNCTIONALLY TESTED FOR ITS PULLBACK SPEEDS. A FAILURE ANALYSIS (FA) LAB TIMER (JMD # (B)(4)) WAS USED DURING THE FUNCTIONAL TEST. TESTS AT BOTH SPEEDS WERE RUN FOR 120 SECONDS TIME DURATION; RESULTS NOTED BELOW. AT 0.5 MM/SEC: EXPECTED PULLBACK DISTANCE: 60 MM. MEASURED PULLBACK DISTANCE: 60 MM. AT 1.0 MM/SEC: EXPECTED PULLBACK DISTANCE: 120 MM. MEASURED PULLBACK DISTANCE: 130 MM. MANUFACTURING ENGINEERING WAS CONTACTED FOR FURTHER PRODUCT ANALYSIS AND COMPLETED THEIR INVESTIGATION ON 07-01-2016. ENGINEERING CONFIRMED THE SPEED ISSUE WITH THE DEVICE AND INDICATED THE CAUSE OF THE FASTER SPEED FAILURE WAS A FAULTY/WORN SPEED SWITCH ON THE UNIT. THIS EVENT IS BEING REPORTED BECAUSE A RATE FASTER THAN THE NOMINAL RATE HAS THE POTENTIAL TO RESULT IN INACCURATE IVUS-BASED LESION LENGTH MEASUREMENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
484988 ENDOSONICS AUTOMATIC PULL BACK DEVICE TRANSDUCER, ULTRASONIC, DIAGNOSTIC ITX ENDOSONICS CORP. R-100

Patients

Seq Age Sex Outcome Treatment
1