ENDOSONICS AUTOMATIC PULL BACK DEVICE
Report
- Report Number
- 2939520-2016-00051
- Event Type
- Malfunction
- Date Received
- July 29, 2016
- Date of Event
- March 8, 2016
- Report Date
- March 8, 2016
- Manufacturer
- ENDOSONICS CORP.
- Product Code
- ITX
- PMA / PMN Number
- K962293
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS CASE WAS REVIEWED AND INVESTIGATED ACCORDING TO THE MANUFACTURER'S POLICY. NO PATIENT INVOLVEMENT; THE ISSUE WAS FOUND DURING A MAINTENANCE CHECK IN THE MANUFACTURER'S WAREHOUSE. THIS IS A REFURBISHED PRODUCT. THE ORIGINAL SERIAL # WAS (B)(4). THERE ARE NO LOT NUMBERS ASSOCIATED WITH THESE DEVICES. THERE IS NO EXPIRATION DATE ASSOCIATED WITH THIS DEVICE. THE FIELD SERVICE ENGINEER TESTED THE DEVICE. NO FACILITY INVOLVEMENT. THIS IS AN EMPLOYEE OF THE MANUFACTURER. THIS IS A REFURBISHED PRODUCT AND DATE NOTED IS DATE IT WAS REFURBISHED. THE ORIGINAL DEVICE MANUFACTURE DATE WAS 03/10/2005. THE DEVICE WILL BE TRANSFERRED TO THE REFURBISH DEPARTMENT TO EVALUATE FOR POSSIBLE REPAIR AND RE-CERTIFICATION. THIS COMPLAINT WILL BE MONITORED AS PART OF COMPLAINT DATA ANALYSIS.
DURING A ROUTINE INSPECTION THE MANUFACTURER'S FIELD SERVICE ENGINEER FOUND THE DEVICE'S PULLBACK SPEED APPEARED FASTER THAN THE STANDARD SPEED SETTING. THE DEVICE WAS REQUESTED TO BE RETURNED FOR ANALYSIS. THE RETURNED DEVICE WAS INVESTIGATED ACCORDING TO THE MANUFACTURER'S POLICY. VISUAL INSPECTION DISCOVERED A MISSING BATTERY COVER AND THE BATTERY ITSELF. NO FURTHER DAMAGE WAS OBSERVED ON THE DEVICE. A KNOWN GOOD 9V BATTERY (MULTIMETER JMD # (B)(4)) WAS INSTALLED INTO THE DEVICE. THE DEVICE WAS FUNCTIONALLY TESTED FOR ITS PULLBACK SPEEDS. A FAILURE ANALYSIS (FA) LAB TIMER (JMD # (B)(4)) WAS USED DURING THE FUNCTIONAL TEST. TESTS AT BOTH SPEEDS WERE RUN FOR 120 SECONDS TIME DURATION; RESULTS NOTED BELOW. AT 0.5 MM/SEC: EXPECTED PULLBACK DISTANCE: 60 MM. MEASURED PULLBACK DISTANCE: 60 MM. AT 1.0 MM/SEC: EXPECTED PULLBACK DISTANCE: 120 MM. MEASURED PULLBACK DISTANCE: 130 MM. MANUFACTURING ENGINEERING WAS CONTACTED FOR FURTHER PRODUCT ANALYSIS AND COMPLETED THEIR INVESTIGATION ON 07-01-2016. ENGINEERING CONFIRMED THE SPEED ISSUE WITH THE DEVICE AND INDICATED THE CAUSE OF THE FASTER SPEED FAILURE WAS A FAULTY/WORN SPEED SWITCH ON THE UNIT. THIS EVENT IS BEING REPORTED BECAUSE A RATE FASTER THAN THE NOMINAL RATE HAS THE POTENTIAL TO RESULT IN INACCURATE IVUS-BASED LESION LENGTH MEASUREMENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 484988 | ENDOSONICS AUTOMATIC PULL BACK DEVICE | TRANSDUCER, ULTRASONIC, DIAGNOSTIC | ITX | ENDOSONICS CORP. | R-100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |