FDA Adverse Event Malfunction Summary report: N

VISTA BASIC - IFVB

MDR report key: 5833104 · Received July 29, 2016

Report

Report Number
9610825-2016-00517
Event Type
Malfunction
Date Received
July 29, 2016
Report Date
July 22, 2016
Manufacturer
B. BRAUN MELSUNGEN AG
Product Code
FRN
PMA / PMN Number
K003029
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2011009. B. BRAUN MEDICAL INC. IS SUBMITTING A SINGLE REPORT ON BEHALF OF B. BRAUN MELSUNGEN (THE MANUFACTURER), AND B. BRAUN MEDICAL INC. (THE IMPORTER). THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INTERNAL REPORT NUMBER (B)(4). THE VOLUMETRIC ACCURACY WAS TESTED THREE TIMES AT 250ML/HR AND 1 HOUR (250ML). THE TEST RESULTS WERE WITHIN SPECIFICATIONS. BASED ON THE RESULTS OF THE INVESTIGATION, THE PUMP OPERATED AS INTENDED. IF ADDITIONAL PERTINENT INFORMATION BECOMES AVAILABLE, A FOLLOW UP WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2011009. B. BRAUN MEDICAL INC. IS SUBMITTING A SINGLE REPORT ON BEHALF OF B. BRAUN MELSUNGEN (THE MANUFACTURER), AND B. BRAUN MEDICAL INC. (THE IMPORTER). THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INTERNAL REPORT NUMBER (B)(4). THE DEVICE INVOLVED HAS BEEN RECEIVED FOR EVALUATION AND THE INVESTIGATION IS ONGOING AT THIS TIME. A FOLLOW UP WILL BE SUBMITTED WHEN THE INVESTIGATION RESULTS BECOME AVAILABLE.

Description of Event or Problem · 0

AS REPORTED BY THE USER FACILITY: CUSTOMER REPORTS NURSE SETS 250 ML TO INFUSE OVER 1 HOUR. PUMP READS INFUSED OVER 1 HOUR, BUT ON ANOTHER CLOCK IT IS INFUSED IN 45 MINUTES. PUMP DOES ALARM AND BAG IS EMPTY. CUSTOMER HAS BEGUN TO SET TIME FOR 1 HOUR AND 10 MINUTES TO HAVE INFUSION CLOSER TO 1 HOUR AS ORDERED. NO PATIENT INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
486026 VISTA BASIC - IFVB INFUSION PUMP FRN B. BRAUN MELSUNGEN AG

Patients

Seq Age Sex Outcome Treatment
1