I-STAT CHEM8+ CARTRIDGE
Report
- Report Number
- 2245578-2016-00071
- Event Type
- Malfunction
- Date Received
- July 28, 2016
- Date of Event
- July 10, 2016
- Report Date
- October 21, 2016
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JGS
- PMA / PMN Number
- K053110
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4)
(B)(4). THE INVESTIGATION WAS COMPLETED ON (B)(6) 2016. RETAIN PRODUCT WAS TESTED AND IS FUNCTIONING ACCORDING TO SPECIFICATION. RETURN PRODUCT WAS NOT AVAILABLE FOR INVESTIGATION.
ON 07/12/2016, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT RESULTS ON A (B)(6) FEMALE PATIENT. THERE WAS NO PATIENT INFORMATION AT THE TIME OF THIS REPORT. NEW INFORMATION WAS RECEIVED ON 07/20/2016 THAT INVOLVED REPEAT TESTING ON I-STAT THAT SUGGESTED A POTENTIAL PRODUCT MALFUNCTION. THE CUSTOMER ALSO STATES THAT RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. THE INVESTIGATION IS UNDERWAY. DATE: (B)(6) 2016, TIME: 15:40, METHOD: I-STAT, LOT: H16038, NA: 118, CI >140; (B)(6) 2016, UNK, I-STAT, H16038, 141, 104, COMMENT: RECEIVED ON 7/20/2016. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. THE INVESTIGATION IS UNDERWAY.
NA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 483513 | I-STAT CHEM8+ CARTRIDGE | CHEM8+ CARTRIDGE | JGS | ABBOTT POINT OF CARE | NA | H16038 |