FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 5824841 · Received July 27, 2016

Report

Report Number
3008382007-2016-37867
Event Type
Malfunction
Date Received
July 27, 2016
Report Date
July 14, 2016
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2016, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING THAT THE SUBJECT METER READ INACCURATELY ERRATIC. THE REPORTER CLAIMED OBTAINING BLOOD GLUCOSE READINGS OF "173, 125 AND 112 MG/DL" WITH THE SUBJECT METER, PERFORMED WITHIN 20 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN'S CRITERIA FOR PRECISION. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
480692 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL

Patients

Seq Age Sex Outcome Treatment
1