FDA Adverse Event Injury Summary report: N

AMISTEM H, HA COATED STEM SIZE 5 STD

MDR report key: 5819155 · Received July 25, 2016

Report

Report Number
3005180920-2016-00379
Event Type
Injury
Date Received
July 25, 2016
Date of Event
June 29, 2016
Report Date
July 25, 2016
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
PMA / PMN Number
K093944
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 20 JULY 2016. LOT 110555: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 05 APRIL 2011. EXPIRATION DATE: 2016-02-29. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. ON 22 JULY 2016 THE MEDICAL AFFAIRS PERFORMED A CLINICAL INVESTIGATION AND COMMENTED AS FOLLOWS: STEM LOOSENING IN A THA ON A LADY OF (B)(6) AFTER 5 YEARS. THE STEM APPEARS TO BE RADIOGRAPHICALLY LOOSE IN THE LATERAL PROXIMAL REGION AND ALSO, TO A LESSER EXTENT, IN THE MEDIAL AREA, THEREFORE REMOVAL IS NECESSARY. NO REASON CAN BE DETERMINED FOR THIS LOOSENING. ASEPTIC LOOSENING IN CEMENTLESS STEMS FOLLOWING THA IS A KNOWN POSSIBLE COMPLICATION THAT IDIOPATHICALLY AFFECTS A SMALL PERCENTAGE OF THE OPERATED JOINTS; A CLEAR CAUSE CANNOT ALWAYS BE FOUND. NOT AVAILABLE.

Description of Event or Problem · 1

REVISION BECAUSE OF STEM LOOSENING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472663 AMISTEM H, HA COATED STEM SIZE 5 STD CEMENTLESS FEMORAL STEM LZO MEDACTA INTERNATIONAL SA 110555

Patients

Seq Age Sex Outcome Treatment
1 64 YR Required Intervention