FDA Adverse Event Injury Summary report: N

DIALOG A + HE/BIC

MDR report key: 5817581 · Received July 22, 2016

Report

Report Number
3002879653-2016-00025
Event Type
Injury
Date Received
July 22, 2016
Date of Event
July 19, 2016
Report Date
July 20, 2016
Manufacturer
B. BRAUN AVITUM AG - MELSUNGEN
Product Code
FKJ
PMA / PMN Number
K083460
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2015012. B. BRAUN MEDICAL INC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF B. BRAUN AVITUM AG (MANUFACTURER). THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN AVITUM INTERNAL REPORT (B)(4). FACILTY STAFF REPORTED THAT DURING DIALYSIS THERAPY THE PATIENT WAS FINE AN THEN SUDDENLY COMPLAINED OF BEING SHORT OF BREATH. PATIENT THEN BECAME UNRESPONSIVE. CPR STARTED BY FACILITY STAFF AND 911 WAS CALLED. BLOOD WAS RETURNED TO THE PATIENT WITH ADDITIONAL NORMAL SALINE. AED WAS APPLIED, WHICH 'NO SHOCK ADVISED', CONTINUE CPR. CPR WAS CONTINUED UNTIL EMS ARRIVED. WHEN EMS ARRIVED, PATIENT WAS DISCONNECTED FROM DIALOG. FACILITY STAFF ALSO REPORTED THAT PATIENT HAD A PULSE WHEN SHE LEFT. PATIENT WAS INTUBATED PRIOR TO BEING TRANSPORTED TO THE HOSPITAL. FACILITY MANAGER WAS UNABLE TO PROVIDE PATIENT MEDICAL INFORMATION. PATIENT CURRENT STATUS IS UNKNOWN BY FACILITY MANAGER. CHIEF BIOMEDICAL TECHNICIAN OBTAINED DIALYSATE SAMPLES WITH CONDUCTIVITY 13.8, PH 7.1. DIALYSATE SAMPLES FOR ELECTROLYTES, COLONY COUNT, AND ENDOTOXINS WERE SENT TO LAB. A FOLLOW UP REPORT WILL BE PROVIDED WHEN THE INVESTIGATION IS COMPLETED.

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2015012. B. BRAUN MEDICAL INC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF B. BRAUN AVITUM AG (MANUFACTURER). THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN AVITUM INTERNAL REPORT (B)(4) CUSTOMER REPORTED THAT A PATIENT TREATED WITH DIALOG+ DIALYSIS MACHINE (SN (B)(6) BECAME UNRESPONSIVE ABOUT 30 MINUTES INTO THERAPY. CARDIOPULMONARY RESUSCITATION (CPR) WAS INITIATED BY THE FACILITY STAFF. THE BLOOD OF THE EXTRACORPOREAL CIRCUIT WAS RETURNED WITH AN ADDITIONAL AMOUNT OF NORMAL SALINE. THE PATIENT WAS TRANSPORTED TO A HOSPITAL BY EMERGENCY MEDICAL SERVICE (EMS). AT THAT TIME THE PATIENT HAD VITAL SIGNS. THE DIALOG+ DIALYSIS MACHINE WAS INSPECTED BY THE FACILITY TECHNICIAN, SUPERVISED BY A B. BRAUN TECHNICIAN. THE DIALOG+ MACHINE OPERATED AS INTENDED. THERE WAS NO MALFUNCTION OR PRODUCT DEVIATION. DIALYSATE SAMPLES WERE TAKEN BY THE USER WHICH IS DONE ROUTINELY AND THE RESULTS WERE: CONDUCTIVITY 13.8 (NORMAL RANGE 13.0 - 15.0) AND PH OF 7.1 (NORMAL RANGE 6.8 - 7.6). DIALYSATE SAMPLES FOR ELECTROLYTES, COLONY COUNT, AND ENDOTOXINS WERE SENT TO A LAB; HOWEVER, THE RESULTS WERE NOT DISCLOSED BY THE FACILITY. AFTER A FOLLOW-UP CALL, THE FACILITY REPORTED THAT THE PATIENT DID NOT PASS AWAY. THE PATIENT WAS TRANSPORTED TO HOSPITAL WITH VITAL SIGNS, AND FROM THE DESCRIPTION THE PATIENT APPEARED TO HAVE HAD A CARDIAC ARREST. DURING THIS CALL, THE FACILITY ALSO STATED THE PATIENT WAS POST HOSPITAL DISCHARGE ONE DAY AND THIS WAS HER FIRST DIALYSIS SINCE DISCHARGE FROM THE HOSPITAL. ALL INFORMATION AND MACHINE TREND FILES ASSOCIATED WITH THIS EVENT WERE FORWARDED TO THE EQUIPMENT MANUFACTURER B. BRAUN AVITUM - GERMANY. ACCORDING TO THEIR INVESTIGATION, THE ANALYSIS OF THE DATA RECORD OF THE DIALOG+ MACHINE SHOWED THAT THE THERAPY WAS STARTED WITH A BLOOD FLOW OF 200 ML/MIN AND THEN INCREASED TO 400 ML/MIN. THEREAFTER THE ALARM "ARTERIAL PRESSURE - LOWER LIMIT" (ALARM CODE 1051) WAS TRIGGERED SEVERAL TIMES AND ACKNOWLEDGED BY THE USER. BECAUSE OF THE ALARMS, THE DIALOG+ MACHINE SWITCHED INTO PATIENT- SAFE MODE, THE BLOOD PUMP STOPPED AND THE MACHINE OPERATED IN BYPASS. THE DATA RECORD SHOWED THAT AFTER THE INCREASE OF THE BLOOD FLOW, THE ARTERIAL PRESSURE IN THE EXTRACORPOREAL CIRCUIT CONTINUOUSLY DECREASED. THIS MIGHT HAVE DIFFERENT REASONS, AS FOR EXAMPLE, A TOO HIGH BLOOD FLOW, AN INADEQUATE POSITIONING OF THE FISTULA NEEDLE OR A REDUCED BLOOD VOLUME DUE TO A HYPOTENSIVE EPISODE. ALL THESE REASONS ARE INDEPENDENT FROM THE MACHINE. AFTER A REDUCTION OF THE BLOOD FLOW TO 350 ML/MIN THE ARTERIAL PRESSURE WAS HIGHER AGAIN AND NO MORE ALARMS OCCURRED UNTIL THE BLOOD PUMP WAS STOPPED BY THE USER. THE BLOOD PUMP STOPPED FOR ABOUT A MINUTE, POSSIBLY DUE TO THE ADMINISTRATION OF ADDITIONAL VOLUME TO THE EXTRACORPOREAL CIRCUIT INDICATED BY AN INCREASE OF THE VENOUS PRESSURE. THE THERAPY WAS CONTINUED FOR ANOTHER 6 MINUTES, UNTIL UF MIN WAS ACTIVATED AND THE PATIENT WAS DISCONNECTED FROM THE MACHINE AFTER ABOUT 24 MINUTES OF THERAPY. NEITHER THE INSPECTION OF THE DIALOG+ MACHINE NOR THE TREND DATA SHOW ANY PRODUCT DEVIATION OR MALFUNCTION. ALL INFORMATION CONCERNING THIS INCIDENT HAS BEEN INCLUDED IN OUR TREND ANALYSIS OF THE PRODUCT LINE. A HISTORICAL REVIEW OF THE CUSTOMER COMPLAINT DATABASE, REVEALED NO ADVERSE TRENDS REGARDING THIS ISSUE. IF ADDITIONAL PERTINENT INFORMATION BECOMES AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

AS REPORTED BY THE USER FACILITY: PER STAFF STATEMENT," PATIENT WAS FINE AN THEN SUDDENLY COMPLAINED OF BEING SHORT OF BREATH". PATIENT THEN BECAME UNRESPONSIVE. AS A FOLLOW UP WITH THE FACILITY, THE B. BRAUN (B)(4) CLINICAL MANAGER LEARNED THAT THE PATIENT HAD JUST BEEN RELEASED FROM THE HOSPITAL AND THE STAFF FELT THAT THE PATIENT WAS UNSTABLE AND HAD QUESTIONS ABOUT THE PATIENT RECEIVING THE TREATMENT. THE FACILITY STAFF ALSO FELT THAT THE MACHINE DID NOT CAUSE THE INCIDENT, BUT THE PATIENT BEING UNSTABLE. THE NURSE MANAGER STATED ABOUT 30 MINUTES INTO THE TREATMENT THE PATIENT STARTED COMPLAINING OF SHORTNESS OF BREATH AND DIFFICULTY BEING ABLE TO BREATH. SHORTLY AFTER THE COMPLAINTS THE PATIENT BECAME UNRESPONSIVE AND CPR WAS STARTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467987 DIALOG A + HE/BIC HAEMODIALYSIS SYSTEM FKJ B. BRAUN AVITUM AG - MELSUNGEN N/A N/A

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| O| R