FDA Adverse Event Death Summary report: N

FUJIFILM

MDR report key: 5816433 · Received July 22, 2016

Report

Report Number
2431293-2016-00029
Event Type
Death
Date Received
July 22, 2016
Report Date
July 1, 2016
Manufacturer
FUJIFILM TECHNO PRODUCTS CO., LTD. MITO SITE
Product Code
FDF
PMA / PMN Number
K090116
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CASE REPORT DESCRIBES THE DEATH OF A PATIENT UNDERGOING A COMPLEX, INTERVENTIONAL BILIARY PROCEDURE THAT USED VARIOUS FUJIFILM PRODUCTS, INCLUDING THE (B)(4). THE (B)(4) IS A CO2 REGULATOR THAT APPEARS TO HAVE BEEN RESPONSIBLE FOR THE INTRODUCTION AND MAINTENANCE OF CO2 GAS LEVELS DURING THE PROCEDURE. THE (B)(4) CO2 REGULATOR IS NOT MARKETED OR SOLD IN THE U.S. HOWEVER, A SIMILAR PRODUCT, THE (B)(4) (K133976) IS MARKETED AND SOLD IN THE U.S. THE DEATH OF THE PATIENT INVOLVED MULTIPLE AND CLEARLY MASS INTRAVASCULAR GAS EMBOLI, APPARENTLY INTRODUCED TO THE PATIENT'S CIRCULATORY SYSTEM VIA THE BILIARY TREE. THE AUTHORS SPECULATE THAT THE PATIENT HAD A PRE-EXISTING BILIOVENOUS SHUNT (A RARE COMMUNICATION BETWEEN THE BILIARY TREE AND THE ADJACENT VEINS), A CONDITION THAT CAN OCCUR WITH LONG-STANDING INFECTION IN THE BILIARY TREE AS WAS THE CASE WITH THIS PATIENT. SIMPLY PUT, THE INFECTION DESTROYS THE TISSUE OF THE BILIARY TREE AND ERODES INTO THE ADJACENT VEINS. IT APPEARS THAT THE CO2 INTRODUCED VIA THE FUJIFILM DEVICE ENTERED THE PATIENT'S CIRCULATION AND ULTIMATELY RESULTED IN THE PATIENT DEATH. GIVEN THE PATIENT DEATH AND THAT THE (B)(4) MAY HAVE CONTRIBUTED TO THE GAS THAT RESULTED IN THAT DEATH, THIS REPORT IS BEING MADE IN AN ABUNDANCE OF CAUTION. IN CONCLUSION, IT APPEARS THAT THE DEVICE PERFORMED AS INTENDED; AND THE PRESUMED CAUSE OF DEATH WAS A CLINICALLY UNRECOGNIZED COMMUNICATION BETWEEN THE BILIARY TREE AND THE PATIENT'S SYSTEMIC CIRCULATION THAT ALLOWED THE GAS EMBOLI TO OCCUR, NOT THE FUJIFILM DEVICE. A TOTAL OF THREE MDR'S WILL BE SUBMITTED (NUMBERED 2431293-2016-00028 THROUGH 2431293-2016-00030) TO INCLUDE THE LISTED CONCOMITANT DEVICES. CUSTOMER UNKNOWN BASED ON ARTICLE.

Description of Event or Problem · 1

AS WRITTEN IN A MEDICAL JOURNAL ARTICLE: A (B)(6) WOMAN WAS ADMITTED TO OUR HOSPITAL WITH A COMPLAINT OF HEPATOLITHIASIS. SHE HAD UNDERGONE ROUX-EN-Y HEPATICOJEJUNOSTOMY FOR CHOLEDOCHAL CYSTS 34 YEARS PREVIOUSLY. WE PERFORMED DPOCS (DIRECT PERORAL CHOLANGIOSCOPY) USING A SHORT-TYPE DOUBLE-BALLOON ENTEROSCOPE (DBE), AN ULTRASLIM ENDOSCOPE, AND AN ENDOSCOPIC CO2 REGULATOR ((B)(4), RESPECTIVELY; FUJIFILM CORP., (B)(4)) WHILE THE PATIENT WAS KEPT ADEQUATELY SEDATED WITH MIDAZOLAM. WE PLANNED TO PERFORM LITHOTRIPSY OF THE HEPATOLITH USING A HOLMIUM: YAG LASER. AFTER WE HAD REACHED THE ANASTOMOSIS USING THE DBE, ATTEMPTS TO EXTRACT THE STONES THROUGH THE DBE USING BALLOON OR BASKET CATHETERS FAILED. WE THEREFORE DECIDED TO PERFORM DPOCS (DIRECT PERORAL CHOLANGIOSCOPY) WITH AN ULTRASLIM ENDOSCOPE PASSED THROUGH AN OVERTUBE USING A PREVIOUSLY DESCRIBED METHOD. THE BALLOON ATTACHED TO THE OVERTUBE REMAINED INFLATED FROM THE TIME THAT WE REACHED THE ANASTOMOSIS UNTIL THE END OF THE PROCEDURE. WE FIRST CONFIRMED THE HEPATOLITH WAS PRESENT. WE THEN PREPARED THE HOLMIUM:YAG LASER FOR LITHOTRIPSY FOR 5 MINUTES, WHILE WE ASPIRATED PUS AND MUCUS DISCHARGED FROM THE PERIPHERAL BILE DUCT NEAR THE HEPATOLITH. AS WE FRACTURED THE HEPATOLITH WITH THE HOLMIUM:YAG LASER, THE PATIENT SUDDENLY WENT INTO SHOCK AND HAD A CARDIAC ARREST. DESPITE IMMEDIATE CARDIOMEGALY RESUSCITATION AND INJECTION OF FLUMAZENIL, SHE DIED. A COMPUTED TOMOGRAPHY (CT) SCAN PERFORMED DURING RESUSCITATION REVEALED MULTIPLE GAS EMBOLI IN THE SYSTEMIC ARTERIES AND VEINS. PATHOLOGICAL EXAMINATION LATER REVEALED HEPATIC ABSCESSES, INFLAMMATION SURROUNDING THE HEPATOLITH, INTRAVASCULAR GAS, AND SYSTEMIC GAS EMBOLI. THERE WAS NO EVIDENCE OF A PATENT FORAMEN OVALE. THE CAUSE OF DEATH WAS SYSTEMIC GAS EMBOLISM. WE BELIEVE ASPIRATION OF PUS AND MUCUS PRIOR TO LITHOTRIPSY MAY HAVE OPENED A PRE-EXISTING BILIOVENOUS SHUNT. SOURCE: TITLE: DEVELOPMENT OF FATAL SYSTEMIC GAS EMBOLISM DURING DIRECT PERORAL CHOLANGIOSCOPY UNDER CARBON DIOXIDE INSUFFLATION AUTHOR: HIROMU KONDO, ITARU NAITOH, TAKAHIRO NAKAZAWA, KAZUKI HAYASHI, YUJI NISHI, SHUICHIRO UMEMURA, TAKASHI JOH; PUBLICATION: ENDOSCOPY; PUBLISHER: THIEME; DATE: JAN 1, 2016. COPYRIGHT© 2016, RIGHTS MANAGED BY (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469834 FUJIFILM DOUBLE BALLOON COLONOSCOPE FDF FUJIFILM TECHNO PRODUCTS CO., LTD. MITO SITE EC-450BI5

Patients

Seq Age Sex Outcome Treatment
1 68 YR Death