ALERE TRIAGE D-DIMER TEST
Report
- Report Number
- 2027969-2016-00530
- Event Type
- Injury
- Date Received
- July 21, 2016
- Date of Event
- June 21, 2016
- Report Date
- June 23, 2016
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- GHH
- PMA / PMN Number
- K042890
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER
Narratives
RETURNED PATIENT SAMPLE YIELDED D-DIMER RESULTS CLOSE TO THE CUSTOMER'S OBSERVED <100 NG/ML WHEN TESTED IN-HOUSE WITH RETAIN LOT W60947. CUSTOMER'S COMPLAINT OF DISCREPANT LOW WAS REPLICATED. CUSTOMER'S COMPLAINT OF DISCREPANT LOW WAS NOT REPLICATED WITH IN-HOUSE TESTING OF RETAIN LOT W60947 AND CARDIAC MARKER CALIBRATOR D. RESULTS WERE FOUND TO BE WITHIN THE PACKAGE INSERT CLAIM. CUSTOMER'S COMPLAINT OF DISCREPANT LOW WAS NOT REPLICATED WITH IN-HOUSE TESTING OF RETAIN LOT W60947 AND D-DIMER ELEVATED DONOR WHOLE BLOOD. RESULTS WERE POSITIVE FOR D-DIMER. DEVICE LOT PERFORMED AS EXPECTED WITH IN-HOUSE SAMPLES, UNABLE TO RULE OUT SAMPLE SPECIFIC INTERFERENCE WAS A POTENTIAL CAUSE FOR DISCREPANT RESULTS. BATCH RECORDS FOR LOT W60947 WERE REVIEWED; LOT MET RELEASE SPECIFICATIONS; NO ISSUES WITH D-DIMER RECOVERY WERE OBSERVED. NO PRODUCT DEFICIENCY WAS ESTABLISHED. NO CORRECTIVE ACTION REQUIRED AT THIS TIME.
ON (B)(6) 2016 A (B)(6) FEMALE CAME IN TO THE URGENT CARE REPORTING REDNESS/PURPLE DISCOLORATION TO GREAT TOE AND CALF. THE TRIAGE D-DIMER TEST RESULT WAS <100NG/ML (RANGE=0.500). THE PATIENT WENT ELSEWHERE TO OBTAIN AN ULTRASOUND WHICH INDICATED THAT THERE WAS A BLOOD CLOT. ON (B)(6) 2016 THE PATIENT RETURNED TO URGENT CARE AND THE TRIAGE D-DIMER TEST WAS REPEATED WITH THE SAME RESULT AS THE PREVIOUS DAY. THE PATIENT'S SAMPLE WAS SENT TO THE HOSPITAL FOR TESTING AND GAVE A RESULT OF 2.07 MCG/ML FEU (RANGE=0-0.5) ON THE STAGO COMPACT ANALYZER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 466940 | ALERE TRIAGE D-DIMER TEST | CARDIAC MARKER TEST | GHH | ALERE SAN DIEGO, INC. | 98100 | W60947B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | Required Intervention |