FDA Adverse Event Injury Summary report: N

OSS-K DIAPHYSEAL SEGMENT 7 CM W/SCREWS

MDR report key: 5810559 · Received July 20, 2016

Report

Report Number
0001825034-2016-02674
Event Type
Injury
Date Received
July 20, 2016
Report Date
June 21, 2016
Manufacturer
BIOMET ORTHOPEDICS
Product Code
JDI
PMA / PMN Number
PK020970
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THE LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 4 STATES, "LOOSENING, MIGRATION, OR FRACTURE OF THE IMPLANTS CAN OCCUR DUE TO LOSS OF FIXATION, TRAUMA, MALALIGNMENT, BONE RESORPTION, EXCESSIVE ACTIVITY." NUMBER 9 STATES, "FATIGUE FRACTURE OF COMPONENT CAN OCCUR AS A RESULT OF LOSS OF FIXATION, STRENUOUS ACTIVITY, MALALIGNMENT, TRAUMA, NON-UNION, OR EXCESSIVE WEIGHT." DEVICE REMAINS IMPLANTED.

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY CORRECTED INFORMATION AND ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH. CONCOMITANT PRODUCTS: OSS FEMORAL BUSHING, CATALOG#: 161034, LOT#: 153220. OSS TIBIAL BEARING, CATALOG#: 150410, LOT#: 464070. OSS FEMORAL MODULAR SEGMENT, CATALOG#: 161012, LOT#: 872880. OSS YOKE, CATALOG#: 133080, LOT#: 133080. OSS AXLE, CATALOG#: 161035, LOT#: 228210. OSS LOCKING PIN, CATALOG#: 150478, LOT#: 158660. OSS STACKING ADAPTER, CATALOG#: 150483, LOT#: 192940. OSS STACKING ADAPTER, CATALOG#: 150483, LOT#: 689000.

Description of Event or Problem · 1

PATIENT HAS BEEN INDICATED FOR REVISION DUE TO A TAPER FRACTURE APPROXIMATELY 7 YEARS POST-IMPLANTATION. NO REVISION HAS BEEN REPORTED TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
462764 OSS-K DIAPHYSEAL SEGMENT 7 CM W/SCREWS PROSTHESIS, HIP JDI BIOMET ORTHOPEDICS N/A 965880

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention