METRIX AIR
Report
- Report Number
- 1052693-2016-01236
- Event Type
- Malfunction
- Date Received
- July 20, 2016
- Date of Event
- June 29, 2016
- Report Date
- July 20, 2016
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- PMA / PMN Number
- K150052
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO
- Reporter Occupation
- OTHER
Narratives
(B)(4). RETURNED METER EVALUATED WITH NO DEFECT FOUND. TEST STRIPS NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER'S EXPECTED NON-FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 80 TO 105MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULTS OF 175 AND 211 MG/DL. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 03/28/2017. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: TRUE METRIX AIR 175 ON (B)(6) 2016 07:00:00 AM FASTING. TRUE METRIXAIR 211 ON (B)(6) 2016 07:00:00 AM FASTING. TRUE METRIXAIR 185 ON (B)(6) 2016 07:00:00 AM FASTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 464051 | METRIX AIR | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | METRIX AIR | MT1499 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |