FDA Adverse Event Malfunction Summary report: N

MCKESSON RADIOLOGY

MDR report key: 5809958 · Received July 20, 2016

Report

Report Number
8022257-2016-00002
Event Type
Malfunction
Date Received
July 20, 2016
Date of Event
April 22, 2016
Report Date
April 22, 2016
Manufacturer
MCKESSON MEDICAL IMAGING COMPANY
Product Code
LLZ
PMA / PMN Number
K140909
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MCKESSON COMPLETED AN INVESTIGATION AND CONCLUDED THAT THE ROOT CAUSE OF THE PROBLEM WAS A SOFTWARE DEFECT. MCKESSON FOUND THAT THE STAT PRIORITY ALERT ICON MAY NOT BE DISPLAYED IN THE LIST, FOR ANY USER, WHEN ALL STAT STUDIES ARE OPENED AT THE SAME TIME AND THERE ARE NO "HIGH" PRIORITY STUDIES IN THE UNREPORTED STUDY LIST. UNDER THESE CONDITIONS, THE STAT PRIORITY BUTTON WILL NOT DISPLAY THE CORRECT NUMBER OF STAT STUDIES. MCKESSON WILL WORK WITH AFFECTED CUSTOMERS TO APPLY A SOFTWARE UPDATE TO PREVENT THE RECURRENCE OF THE PROBLEM.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE STAT PRIORITY ICON WAS NOT DISPLAYED IN THE STUDY LIST WHEN THE STUDY WAS OPENED. NO PATIENT HARM WAS REPORTED DUE TO THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
464238 MCKESSON RADIOLOGY PICTURE ARCHIVING AND COMMUNICATION SYSTEM LLZ MCKESSON MEDICAL IMAGING COMPANY 12.1.1