FDA Adverse Event
Malfunction
Summary report: N
MCKESSON RADIOLOGY
MDR report key: 5809958
·
Received July 20, 2016
Report
- Report Number
- 8022257-2016-00002
- Event Type
- Malfunction
- Date Received
- July 20, 2016
- Date of Event
- April 22, 2016
- Report Date
- April 22, 2016
- Manufacturer
- MCKESSON MEDICAL IMAGING COMPANY
- Product Code
- LLZ
- PMA / PMN Number
- K140909
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
MCKESSON COMPLETED AN INVESTIGATION AND CONCLUDED THAT THE ROOT CAUSE OF THE PROBLEM WAS A SOFTWARE DEFECT. MCKESSON FOUND THAT THE STAT PRIORITY ALERT ICON MAY NOT BE DISPLAYED IN THE LIST, FOR ANY USER, WHEN ALL STAT STUDIES ARE OPENED AT THE SAME TIME AND THERE ARE NO "HIGH" PRIORITY STUDIES IN THE UNREPORTED STUDY LIST. UNDER THESE CONDITIONS, THE STAT PRIORITY BUTTON WILL NOT DISPLAY THE CORRECT NUMBER OF STAT STUDIES. MCKESSON WILL WORK WITH AFFECTED CUSTOMERS TO APPLY A SOFTWARE UPDATE TO PREVENT THE RECURRENCE OF THE PROBLEM.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE STAT PRIORITY ICON WAS NOT DISPLAYED IN THE STUDY LIST WHEN THE STUDY WAS OPENED. NO PATIENT HARM WAS REPORTED DUE TO THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 464238 | MCKESSON RADIOLOGY | PICTURE ARCHIVING AND COMMUNICATION SYSTEM | LLZ | MCKESSON MEDICAL IMAGING COMPANY | 12.1.1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |