FDA Adverse Event Other Summary report: N

BD PARADIGM LINK BLOOD GLUCOSE MONITOR

MDR report key: 580878 · Received March 1, 2005

Report

Report Number
2243072-2005-00064
Event Type
Other
Date Received
March 1, 2005
Date of Event
February 2, 2005
Report Date
February 28, 2005
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

CONSUMER CALLED AGAIN TO REPORT GETTING ERRATIC READINGS WITHIN A SHORT PERIOD OF TIME. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVG READING (190) AS REF. POINT VS. BD READINGS OF 22, 230, 227, 281 YIELDED DATA POINTS IN THE E ZONE OF THE GRID, OUTSIDE OF ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD PARADIGM LINK BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 4064225

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other