BD PARADIGM LINK BLOOD GLUCOSE MONITOR
MDR report key: 580878
·
Received March 1, 2005
Report
- Report Number
- 2243072-2005-00064
- Event Type
- Other
- Date Received
- March 1, 2005
- Date of Event
- February 2, 2005
- Report Date
- February 28, 2005
- Manufacturer
- BECTON DICKINSON
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
CONSUMER CALLED AGAIN TO REPORT GETTING ERRATIC READINGS WITHIN A SHORT PERIOD OF TIME. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVG READING (190) AS REF. POINT VS. BD READINGS OF 22, 230, 227, 281 YIELDED DATA POINTS IN THE E ZONE OF THE GRID, OUTSIDE OF ACCEPTABLE LIMITS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD PARADIGM LINK BLOOD GLUCOSE MONITOR | BD TEST STRIPS | NBW | BECTON DICKINSON | NA | 4064225 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |