MODULAR CAPTURE - DISTAL RESECTION
Report
- Report Number
- 0002249697-2016-02384
- Event Type
- Malfunction
- Date Received
- July 20, 2016
- Date of Event
- June 16, 2016
- Report Date
- June 23, 2016
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- MBH
- PMA / PMN Number
- K123486
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
WHEN COMPLETED, THE INVESTIGATION RESULTS WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.
AN EVENT REGARDING SEATING/LOCKING ISSUE INVOLVING A MODULAR CAPTURE ¿ DISTAL RESECTION WAS REPORTED. THE EVENT WAS NOT CONFIRMED. METHOD AND RESULTS: DEVICE EVALUATION AND RESULTS: DURING THE FUNCTION INSPECTION THE DISTAL RESECTION GUIDE WAS ABLE TO SNAP IN WITH THE MODULAR CAPTURE. THE DISTAL RESECTION GUIDE HELD SECURELY ONTO THE MODULAR CAPTURE AS INTENDED WITH NO SEPARATION TRYING TO PULL APART. MEDICAL RECORDS RECEIVED AND EVALUATION: NOT PERFORMED AS NO MEDICAL RECORDS WERE RETURNED FOR EVALUATION. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS INDICATES DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. COMPLAINT HISTORY REVIEW: THERE HAVE BEEN NO OTHER EVENTS FOR THIS LOT. CONCLUSIONS: THE EXACT CAUSE OF THE EVENT COULD NOT BE DETERMINED BECAUSE DURING THE FUNCTIONAL INSPECTION THE DISTAL RESECTION GUIDE WAS ABLE TO SNAP IN WITH THE MODULAR CAPTURE. THE DISTAL RESECTION GUIDE HELD SECURELY ONTO THE MODULAR CAPTURE AS INTENDED WITH NO SEPARATION TRYING TO PULL APART. THE EVENT COULD NOT BE REPLICATED SINCE THE DISTAL RESECTION GUIDE SNAPPED INTO THE MODULAR CAPTURE AS INTENDED.
THE CAPTURE WOULD NOT LOCK INTO THE CUTTING BLOCK. USED WITHOUT CAPTURE.
THE CAPTURE WOULD NOT LOCK INTO THE CUTTING BLOCK. USED WITHOUT CAPTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 462604 | MODULAR CAPTURE - DISTAL RESECTION | PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO | MBH | STRYKER ORTHOPAEDICS-MAHWAH | 15ER9TF1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |